Accurate in vivo measurement of polyethylene wear in total knee arthroplasty

被引:18
作者
Hoshino, A
Fukuoka, Y
Ishida, A
机构
[1] Kawaguchi Kohgyo Gen Hosp, Dept Orthopaed Surg, Kawaguchi, Saitama 3320031, Japan
[2] Tokyo Med & Dent Univ, Inst Biomat & Bioengn, Tokyo, Japan
关键词
total knee arthroplasty (TKA); polyethylene wear; computer vision; three-dimensional (3D)/two-dimensional (2D) matching;
D O I
10.1054/arth.2002.32172
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This study describes a new technique to evaluate accurate linear polyethylene wear and reports the in vivo annual linear wear rate in total knee arthroplasty. Patients (n = 69) were assessed using a single standing anteroposterior radiograph with 8 reference markers around the knee, where 3-dimensional OD) locations of markers and 3D geometry of the implant were known. By matching the 3D locations of the markers and their projected points on the radiograph, 3D/2D transformation parameters were estimated. Using these parameters, the 3D coordinates of the implant were estimated from the projected x-ray silhouette. After calculating the temorotibial shortest distance, wear and creep deformity of the polyethylene was defined with all accuracy of 0.1 mm or better. Measured linear wear in the 69 knees varied from 0 mm (1 month postoperatively) to 2.71 min (7 years postoperatively). The correlation of linear wear (Y) and year (X) was Y = 0.23 X -0.11 (P<.001). This Study revealed that annual wear rate was 0.23 mm and 2 times greater than the average annual wear rate in total hip arthroplasties.
引用
收藏
页码:490 / 496
页数:7
相关论文
共 26 条
[1]  
Abdel-Aziz Y.I., 1971, P S CLOS RANG PHOT, V18, P103, DOI [10.14358/PERS.81.2.103, DOI 10.1080/10671188.1967.10616517, DOI 10.14358/PERS.81.2.103]
[2]   POLYETHYLENE WEAR IN MENISCAL KNEE REPLACEMENT - A ONE TO 9-YEAR RETRIEVAL ANALYSIS OF THE OXFORD KNEE [J].
ARGENSON, JN ;
OCONNOR, JJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (02) :228-232
[3]   In vivo kinematics of cruciate-retaining and -substituting knee arthroplasties [J].
Banks, SA ;
Markovich, GD ;
Hodge, WA .
JOURNAL OF ARTHROPLASTY, 1997, 12 (03) :297-304
[4]   Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy [J].
Banks, SA ;
Hodge, WA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (06) :638-649
[5]   The incidence of modular tibial polyethylene insert exchange in total knee arthroplasty when polyethylene failure occurs [J].
Bert, JM ;
Reuben, J ;
Kelly, F ;
Gross, M ;
Elting, J .
JOURNAL OF ARTHROPLASTY, 1998, 13 (06) :609-614
[6]   Wear in retrieved condylar knee arthroplasties - A comparison of wear in different designs of 280 retrieved condylar knee prostheses [J].
Blunn, GW ;
Joshi, AB ;
Minns, RJ ;
Lidgren, L ;
Lilley, P ;
Ryd, L ;
Engelbrecht, E ;
Walker, PS .
JOURNAL OF ARTHROPLASTY, 1997, 12 (03) :281-290
[7]  
COLLIER JP, 1991, CLIN ORTHOP RELAT R, P232
[8]  
Dennis DA, 1996, CLIN ORTHOP RELAT R, P107
[9]  
Fukuoka Y, 1999, IEEE Trans Rehabil Eng, V7, P228, DOI 10.1109/86.769413
[10]   RETRIEVAL ANALYSIS OF TOTAL KNEE PROSTHESES - A METHOD AND ITS APPLICATION TO 48 TOTAL CONDYLAR PROSTHESES [J].
HOOD, RW ;
WRIGHT, TM ;
BURSTEIN, AH .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (05) :829-842