A multielectrode implant device for the cerebral cortex

被引:64
作者
deCharms, RC
Blake, DT [1 ]
Merzenich, MM
机构
[1] Univ Calif San Francisco, Coleman Lab, San Francisco, CA 94143 USA
[2] Univ Calif San Francisco, Keck Ctr Integrat Neurosci, San Francisco, CA 94143 USA
关键词
array; auditory; behavior; population coding; primate; somatosensory;
D O I
10.1016/S0165-0270(99)00087-4
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A new class of brain implant technology was developed that allows the simultaneous recording of voltage signals from many individual neurons in the cerebral cortex during cognitive tasks. The device allows recording from 49 independent positions spanning a 2 x 2-mm region of neural tissue. The recording electrodes are positioned in a square grid with 350 mu m spacing, and each microelectrode can be precisely independently vertically positioned using a hydraulic microdrive. The device utilizes ultrafine, sharp iridium microelectrodes that minimize mechanical disturbance of the region near the electrode tip and produce low noise neuronal recordings. The total weight of this device is less than 20 g, and the device is reusable. The implant device has been used for transdural recordings in primary somatosensory and auditory cortices of marmosets, owl monkeys, and rats. On a typical day, one-third of the microelectrodes yield well-discriminated single neuron action potential waveforms. Additional array electrodes yield lower amplitude driven multiunit activity. The average signal-to-noise ratio of discriminated action potential waveforms 6 months after implantation was greater than 9. Simple design alternatives are discussed that can increase the number of electrodes in the array and the depths at which dense array recordings can be achieved. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:27 / 35
页数:9
相关论文
共 20 条