Computer-assisted total knee replacement - A controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT protocol)

被引:214
作者
Chauhan, SK
Clark, GW
Lloyd, S
Scott, RG
Breidahl, W
Sikorski, JM
机构
[1] Perth Radiol Clin, Perth, WA 6008, Australia
[2] Hollywood Private Hosp, Perth, WA, Australia
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2004年 / 86B卷 / 06期
关键词
D O I
10.1302/0301-620X.86B6.15456
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A controlled study, comparing computer- and conventional jig-assisted total knee replacement in six cadavers is presented. In order to provide a quantitative assessment of the alignment of the replacements, a CT-based technique which measures seven parameters of alignment has been devised and used. In this a multi-slice CT machine scanned in 2.5 mm slices from the acetabular roof to the dome of the talus with the subject's legs held in a standard position. The mechanical and anatomical axes were identified, from three-dimensional landmarks, in both anteroposterior and lateral planes. The coronal and sagittal alignment of the prosthesis was then measured against the axes. The rotation of the femoral component was measured relative to the transepicondylar axis. The rotation of the tibial component was measured with reference to the posterior tibial condyles and the tibial tuberosity. Coupled femorotibial rotational alignment was assessed by superimposition of the femoral and tibial axial images. The radiation dose was 2.7 mSV. The computer-assisted total knee replacements showed better alignment in rotation and flexion of the femoral component, the posterior slope of the tibial component and in the matching of the femoral and tibial components in rotation. Differences were statistically significant and of a magnitude that support extension of computer assistance to the clinical situation.
引用
收藏
页码:818 / 823
页数:6
相关论文
共 20 条
[1]   FEMORAL ROTATIONAL ALIGNMENT, BASED ON THE ANTEROPOSTERIOR AXIS, IN TOTAL KNEE ARTHROPLASTY IN A VALGUS KNEE - A TECHNICAL NOTE [J].
ARIMA, J ;
WHITESIDE, LA ;
MCCARTHY, DS ;
WHITE, SE .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77 (09) :1331-1334
[2]  
BARGREN JH, 1983, CLIN ORTHOP RELAT R, P178
[3]  
Barrack RL, 2001, CLIN ORTHOP RELAT R, P46
[4]  
BERGER RA, 1993, CLIN ORTHOP RELAT R, P40
[5]   Malrotation causing patellofemoral complications after total knee arthroplasty [J].
Berger, RA ;
Crossett, LS ;
Jacobs, JJ ;
Rubash, HE .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (356) :144-153
[6]  
ECKHOFF DG, 1995, CLIN ORTHOP RELAT R, P176
[7]   THE EFFECT OF ALIGNMENT OF THE IMPLANT ON FRACTURES OF THE PATELLA AFTER CONDYLAR TOTAL KNEE ARTHROPLASTY [J].
FIGGIE, HE ;
GOLDBERG, VM ;
FIGGIE, MP ;
INGLIS, AE ;
KELLY, M ;
SOBEL, M .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (07) :1031-1039
[8]   A computer model to simulate patellar biomechanics following total knee replacement: the effects of femoral component alignment [J].
Heegaard, JH ;
Leyvraz, PF ;
Hovey, CB .
CLINICAL BIOMECHANICS, 2001, 16 (05) :415-423
[9]  
HOOD RW, 1981, CLIN ORTHOP RELAT R, P94
[10]   CORONAL ALIGNMENT AFTER TOTAL KNEE REPLACEMENT [J].
JEFFERY, RS ;
MORRIS, RW ;
DENHAM, RA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (05) :709-714