The mechanical effect of blocking screws ("Poller screws") in stabilizing tibia fractures with short proximal or distal fragments after insertion of small-diameter intramedullary nails

被引:254
作者
Krettek, C
Miclau, T
Schandelmaier, F
Stephan, C
Möhlmann, U
Tscherne, H
机构
[1] Monash Univ, Alfred Hosp, Dept Surg, Melbourne, Vic 3181, Australia
[2] Univ Calif San Francisco, San Francisco Gen Hosp, Dept Orthopaed Surg, San Francisco, CA USA
[3] Hannover Med Sch, Trauma Dept, D-3000 Hannover, Germany
关键词
blocking screw; small-diameter intramedullary nail; tibia; metaphyseal fracture;
D O I
10.1097/00005131-199911000-00006
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives/Hypothesis: To evaluate the mechanical effects of medial and lateral blocking screws in supplementing intramedullary nail fixation of high proximal and low distal tibial fractures treated with small-diameter intramedullary nails. Study Design: Intact fresh human cadaveric tibiae were sectioned to provide ten distal segments measuring seventy millimeters and ten proximal segments measuring ninety millimeters. In the distal segments, stainless steel solid eight-millimeter tibial nails were advanced to eight millimeters from the ankle joint. Two transverse and one anterior-posterior (AP) locking screw were inserted using a custom-made jig. The same jig was used for the placement of a medial and a lateral blocking screw (BS) in the AP direction, nine millimeters above the superiormost interlocking screw and eight millimeters distal to the lower end of the segment. In the proximal segments, two interlocking screws (both static and dynamic screws) were placed in a medial-lateral direction with the use of the insertion handle. A jig was used for placement of a medial and a lateral BS in the AP direction, nine millimeters below the lower transverse interlocking screw and sixteen millimeters proximal to the lower end of the segment. The bone-implant construct (BIC) was embedded and fixed in a materials testing machine. The BICs were loaded in the medial-lateral direction at a distance of 185 millimeters from the nail ends with loads from -150 newtons to +150 newtons. Force-displacement curves were recorded before and after insertion of the BSs. Results: In proximal BICs, the addition of BSs decreased the deformation of the BICs 25 percent, from 8.9 +/- 1.9 degrees [mean +/- standard deviation (SD)] in the control group to 6.8 +/- 1.1 degrees in the BS group (mean +/- SD) (p < 0.0001). In distal BICs, the addition of BSs decreased the deformation of the BICs 57 percent, from 9.5 +/- 1.4 degrees (mean +/- SD) in the control group to 4.0 +/- 1.0 degrees in the BS group (mean +/- SD) (p < 0.0001). Conclusions: The study suggests that medial and lateral blocking screws can increase the primary stability of distal and proximal metaphyseal fractures after nailing and can be an effective tool for selected cases that exhibit malalignment and/or instability.
引用
收藏
页码:550 / 553
页数:4
相关论文
共 10 条
[1]
Ahlers J, 1992, Unfallchirurgie, V18, P31, DOI 10.1007/BF02588237
[2]
DONALD G, 1983, CLIN ORTHOP RELAT R, P64
[3]
FREEDMAN EL, 1995, CLIN ORTHOP RELAT R, P25
[4]
INFLUENCES OF SOME DESIGN PARAMETERS ON THE BIOMECHANICS OF THE UNREAMED TIBIAL INTRAMEDULLARY NAIL [J].
HENLEY, MB ;
MEIER, M ;
TENCER, AF .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1993, 7 (04) :311-319
[5]
KRETTEK C, 1995, CLIN ORTHOP RELAT R, P34
[6]
Unreamed intramedullary nailing of femoral shaft fractures: Operative technique and early clinical experience with the standard locking option [J].
Krettek, C ;
Rudolf, J ;
Schandelmaier, P ;
Guy, P ;
Konemann, B ;
Tscherne, H .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1996, 27 (04) :233-254
[7]
KRETTEK C, 1994, UNFALLCHIRURG, V97, P575
[8]
KRETTEK C, 1994, UNFALLCHIRURG, V97, P549
[9]
KRETTEK C, 1999, J BONE JOINT SURG B, V81
[10]
Tornetta P, 1996, CLIN ORTHOP RELAT R, P185