Sensitivity estimation of CMOS optical BDJ detector

被引:14
作者
Lu, GN
Galvan, JM
Jeloyan, C
Goumy, G
Marcoux, V
机构
[1] Lab Imagerie & Signal Acoust CPE, F-69616 Villeurbanne, France
[2] Univ Lyon 1, Lab Elect Nantechnol Capteurs, F-69622 Villeurbanne, France
来源
MATERIALS SCIENCE & ENGINEERING C-BIOMIMETIC AND SUPRAMOLECULAR SYSTEMS | 2002年 / 21卷 / 1-2期
关键词
sensitivity; CMOS BDJ detector; low-noise preamplifiers; synchronous detection;
D O I
10.1016/S0928-4931(02)00087-5
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
We report the characterisation of the CMOS optical Buried Double p-n Junction (BDJ) detector in terms of sensitivity, with the aim of developing micro-systems for biochemical analysis. We have first used a statistical approach with pA-meters for signal and noise measurements. The obtained results show a minimum noise level limiting the sensitivity performance, due to noise contributions of the coupled measuring instruments. Then we have adopted another approach with implementation of noise-reduction techniques, including low-noise preamplification and synchronous detection. This has allowed the sensitivity estimation much closer to the achievable performance of the detector. The CMOS BDJ detector has two outputs, and its detectivity depends on the used output signals. When the detector is employed as a photodetector with its output current (I-1+I-2) as response, the evaluated detectivity is 2.4x10(12) cm Hz(1/2) W-1. When choosing the other output signal I-1, or using both outputs for wavelength detection, the corresponding detectivity is 1.9x10(12) cm Hz(1/2) W-1. By using the synchronous detection technique, an optical signal of less than 10 fW (10(-14) W) is still detectable. (C) 2002 Published by Elsevier Science B.V.
引用
收藏
页码:203 / 210
页数:8
相关论文
共 10 条
[1]   Investigation of CMOS BDJ detector for fluorescence detection in microarray analysis [J].
Lu, GN ;
Guillaud, G ;
Sou, G ;
Devigny, F ;
Pitaval, M ;
Morin, P .
1ST ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS, 2000, :381-386
[2]   A dual-wavelength method using the BDJ detector and its application to iron concentration measurement [J].
Lu, GN .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1999, 10 (04) :312-315
[3]   Design and testing of a CMOS BDJ detector for integrated micro-analysis systems [J].
Lu, GN ;
Sou, G ;
Devigny, F ;
Guillaud, G .
MICROELECTRONICS JOURNAL, 2001, 32 (03) :227-234
[4]   Colour detection using a buried double p-n junction structure implemented in the CMOS process [J].
Lu, GN ;
Chouikha, MB ;
Sou, G ;
Sedjil, M .
ELECTRONICS LETTERS, 1996, 32 (06) :594-596
[5]  
Meade M.L., 1983, LOCK AMPLIFIERS PRIN
[6]  
MOTCHENBACER CD, 1993, LOW NOISE ELECT SYST, pCH8
[7]   Integrated microsystem for blue/UV detection [J].
Pauchard, A ;
Furrer, B ;
Randjelovic, Z ;
Rochas, A ;
Manic, D ;
Popovic, RS .
SENSORS AND ACTUATORS A-PHYSICAL, 2000, 85 (1-3) :99-105
[8]   A seawater pH determination method using a BDJ detector [J].
Sedjil, W ;
Lu, GN .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1998, 9 (09) :1587-1592
[9]   An integrated CMOS microluminometer for low-level luminescence sensing in the bioluminescent bioreporter integrated circuit [J].
Simpson, ML ;
Sayler, GS ;
Patterson, G ;
Nivens, DE ;
Bolton, EK ;
Rochelle, JM ;
Arnott, JC ;
Applegate, BM ;
Ripp, S ;
Guillorn, MA .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 72 (02) :134-140
[10]  
VANVLIET KM, 1965, FLUCTUATION PHENOMEN, pCH7