Immobilization of acetylcholinesterase on screen-printed electrodes: comparative study between three immobilization methods and applications to the detection of organophosphorus insecticides
被引:190
作者:
Andreescu, S
论文数: 0引用数: 0
h-index: 0
机构:Univ Perpignan, Ctr Phytopharm, URA 5054, F-66860 Perpignan, France
Andreescu, S
Barthelmebs, L
论文数: 0引用数: 0
h-index: 0
机构:Univ Perpignan, Ctr Phytopharm, URA 5054, F-66860 Perpignan, France
Barthelmebs, L
Marty, JL
论文数: 0引用数: 0
h-index: 0
机构:Univ Perpignan, Ctr Phytopharm, URA 5054, F-66860 Perpignan, France
Marty, JL
机构:
[1] Univ Perpignan, Ctr Phytopharm, URA 5054, F-66860 Perpignan, France
[2] Univ Bucharest, Fac Chem, Dept Analyt Chem, Bucharest 76234, Romania
The analytical performance of three acetylcholinesterase (AChE) screen-printed biosensors, designed for the detection of pesticides are evaluated. Bioencapsulation of the enzyme in a sol-gel composite and immobilization by metal-chelate affinity were compared with the entrapment of the enzyme in a photopolymerisable polymer. A very low amount of enzyme ranging between 0.8 and 1.2 mIU was immobilized on the electrode surface in each approach. The sensors exhibited a storage stability of over 6 months when the enzyme was encapsulated in a polymer film. Pesticide concentrations in the range of 10(-8) to 10(-9) M paraoxon, dichlorvos and chlorpyrifos ethyl oxon could be detected according to each configuration by following an incubation time of 20 min. (C) 2002 Elsevier Science B.V. All rights reserved.