Optimization and Validation of a Multiplex, Electrochemiluminescence-Based Detection Assay for the Quantitation of Immunoglobulin G Serotype-Specific Antipneumococcal Antibodies in Human Serum

被引:72
作者
Marchese, Rocio D. [1 ]
Puchalski, Derek [1 ]
Miller, Pamela [2 ]
Antonello, Joseph [2 ]
Hammond, Olivia [1 ]
Green, Tina [2 ]
Rubinstein, Leonard J. [3 ]
Caulfield, Michael J. [1 ]
Sikkema, Daniel [1 ]
机构
[1] Merck Res Labs, Wayne Clin Support, Wayne, PA USA
[2] Merck Res Labs, Nonclin Stat, West Point, PA USA
[3] Merck Res Labs, ID Vaccines, Upper Gwynedd, PA USA
关键词
LINKED-IMMUNOSORBENT-ASSAY; STANDARD REFERENCE SERUM; PNEUMOCOCCAL POLYSACCHARIDES; VACCINE; IMMUNOASSAYS; ASSIGNMENT; CHILDREN; DISEASE; UNITS;
D O I
10.1128/CVI.00415-08
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
Pneumovax 23 consists of a mixture of highly purified capsular polysaccharides (Ps) from 23 of the most prevalent serotypes of Streptococcus pneumoniae. Testing of vaccine immunogenicity has been historically performed on the enzyme-linked immunosorbent assay (ELISA) platform, validated to measure immunoglobulin G (IgG) antibodies to all 23 serotypes included in Pneumovax 23. In order to significantly improve the throughput of this form of testing, we have developed and validated a direct binding electrochemiluminescence (ECL)-based multiplex assay that can measure the antibody response in human serum to eight serotypes within a single microtiter well. The pneumococcal (Pn) ECL assay is based on the Meso Scale Discovery (MSD) technology which utilizes a Sulfo-Tag-labeled anti-human IgG antibody that emits light upon electrochemical stimulation. The Pn ECL assay exhibits a wide dynamic range and provides the ability to read concentrations down to the minimum reported concentration in the Merck ELISA (0.1 mu g/ml). Cross-reactivity assessment using type-specific monoclonal antibodies showed no cross talk between antigen spots within a well. By use of the WHO Pn sample reference panel, the results obtained with the Pn ECL assay were compared to the results obtained with the international Pn ELISA. The results for the Pn ECL assay satisfied the WHO-recommended acceptance criterion for concordance for all seven serotypes with published Pn ELISA values, and the overall correlation (r value) across the seven serotypes was 0.994. The MSD methodology has great potential to be extremely useful for simultaneously quantitating IgG responses to several Pn serotypes while conserving serum volumes and laboratory testing time.
引用
收藏
页码:387 / 396
页数:10
相关论文
共 28 条
[1]  
[Anonymous], 2000, MMWR Recomm Rep, V49, P1
[2]  
[Anonymous], 2001, GUID IND BIOAN METH
[3]   Development and comparison of two immunoassay formats for rapid detection of botulinum neurotoxin type A [J].
Attree, Olivier ;
Guglielmo-Viret, Valerie ;
Gros, Valerie ;
Thullier, Philippe .
JOURNAL OF IMMUNOLOGICAL METHODS, 2007, 325 (1-2) :78-87
[4]  
BLACKBURN GF, 1991, CLIN CHEM, V37, P1534
[5]  
*COMM ISS PRIOR NE, 1986, NEW VACC DEV EST PRI, V2, P357
[6]   A NEW NONISOTOPIC DETECTION SYSTEM FOR IMMUNOASSAYS [J].
DEAVER, DR .
NATURE, 1995, 377 (6551) :758-760
[7]   The burden of pneumococcal disease among adults in developed and developing countries: what is and is not known [J].
Fedson, DS ;
Anthony, J ;
Scott, G .
VACCINE, 1999, 17 :S11-S18
[8]   Validation of immunoassays for bioanalysis: a pharmaceutical industry perspective [J].
Findlay, JWA ;
Smith, WC ;
Lee, JW ;
Nordblom, GD ;
Das, I ;
DeSilva, BS ;
Khan, MN ;
Bowsher, RR .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2000, 21 (06) :1249-1273
[9]   The epidemiology of pneumococcal infection in children in the developing world [J].
Greenwood, B .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY B-BIOLOGICAL SCIENCES, 1999, 354 (1384) :777-785
[10]  
HAALAND PD, 1989, EXPT DESIGN BIOTECNO, P37