Monte Carlo calculations of the total-to-peak ratio in gamma-ray spectrometry

被引:9
作者
Korun, M [1 ]
Vidmar, T [1 ]
机构
[1] Jozef Stefan Inst, Ljubljana 1000, Slovenia
关键词
Electron scattering - Mathematical models - Monte Carlo methods - Probability - Radiation detectors - Radioactivity measurement - Spectrum analysis;
D O I
10.1016/S0969-8043(99)00245-6
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Monte Carlo calculations of the total-to-peak ratio for a coaxial closed end n-type gamma-ray detector were performed and the results compared to the measured values. Whereas reasonable agreement between the calculated and measured spatial dependencies was achieved, the calculation overestimates the absolute experimental values by about 10%. The spatial dependence of the total-to-peak ratio is discussed in terms of detector geometry. (C) 2000 Elsevier Science Ltd. All rights reserved.