Quantitative assessment of selective in-plane shielding of tissues in computed tomography through evaluation of absorbed dose and image quality

被引:100
作者
Geleijns, J.
Salvado Artells, M.
Veldkamp, W. J. H.
Lopez Tortosa, M.
Calzado Cantera, A.
机构
[1] Leiden Univ, Med Ctr, Dept Radiol, NL-2333 ZA Leiden, Netherlands
[2] Univ Rovira & Virgili, Dept Ciencies Med Basiques, Fac Med & Ciencies Salut, Tarragona 43201, Spain
[3] Univ Complutense Madrid, Dept Radiol, E-28040 Madrid, Spain
关键词
X-ray computed tomography; radiation protection; radiation effects;
D O I
10.1007/s00330-006-0217-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study aimed at assessment of efficacy of selective in-plane shielding in adults by quantitative evaluation of the achieved dose reduction and image quality. Commercially available accessories for in-plane shielding of the eye lens, thyroid and breast, and an anthropomorphic phantom were used for the evaluation of absorbed dose and image quality. Organ dose and total energy imparted were assessed by means of a Monte Carlo technique taking into account tube voltage, tube current, and scanner type. Image quality was quantified as noise in soft tissue. Application of the lens shield reduced dose to the lens by 27% and to the brain by 1%. The thyroid shield reduced thyroid dose by 26%; the breast shield reduced dose to the breasts by 30% and to the lungs by 15%. Total energy imparted (unshielded/shielded) was 88/86 mJ for computed tomography (CT) brain, 64/60 mJ for CT cervical spine, and 289/260 mJ for CT chest scanning. An increase in image noise could be observed in the ranges were bismuth shielding was applied. The observed reduction of organ dose and total energy imparted could be achieved more efficiently by a reduction of tube current. The application of in-plane selective shielding is therefore discouraged.
引用
收藏
页码:2334 / 2340
页数:7
相关论文
共 19 条
[1]  
[Anonymous], 1991, ANN ICRP, V60
[2]   Would thyroid and breast shielding be beneficial in CT of the head? [J].
Beaconsfield, T ;
Nicholson, R ;
Thornton, A ;
Al-Kutoubi, A .
EUROPEAN RADIOLOGY, 1998, 8 (04) :664-667
[3]   Efficacy of breast shielding during CT of the head [J].
Brnic, Z ;
Vekic, B ;
Hebrang, A ;
Anic, P .
EUROPEAN RADIOLOGY, 2003, 13 (11) :2436-2440
[4]  
Ford R., 1978, SLAC210
[5]   In-plane bismuth breast shields for pediatric CT: Effects on radiation dose and image quality using experimental and clinical data [J].
Fricke, BL ;
Donnelly, LF ;
Frush, DP ;
Yoshizumi, T ;
Varchena, V ;
Poe, SA ;
Lucaya, J .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 180 (02) :407-411
[6]   Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions [J].
Greess, H ;
Wolf, H ;
Baum, U ;
Lell, M ;
Pirkl, M ;
Kalender, W ;
Bautz, WA .
EUROPEAN RADIOLOGY, 2000, 10 (02) :391-394
[7]   Low-dose CT of the paranasal sinuses with eye lens protection: effect on image quality and radiation dose [J].
Hein, E ;
Rogalla, P ;
Klingebiel, R ;
Hamm, B .
EUROPEAN RADIOLOGY, 2002, 12 (07) :1693-1696
[8]   Orbital, thyroid, and breast superficial radiation shielding for patients undergoing diagnostic CT [J].
Hopper, KD .
SEMINARS IN ULTRASOUND CT AND MRI, 2002, 23 (05) :423-427
[9]  
Hopper KD, 2001, AM J NEURORADIOL, V22, P1194
[10]   The breast: In-plane x-ray protection during diagnostic thoracic CT - Shielding with bismuth radioprotective garments [J].
Hopper, KD ;
King, SH ;
Lobell, ME ;
TenHave, TR ;
Weaver, JS .
RADIOLOGY, 1997, 205 (03) :853-858