A survey of heparin monitoring in Australasia

被引:7
作者
MacKinlay, N
Favaloro, E
Arthur, C
Smith, J
Aboud, M
机构
[1] WESTMEAD HOSP, INST CLIN PATHOL & MED RES, DEPT HEMATOL, WESTMEAD, NSW, AUSTRALIA
[2] WESTMEAD HOSP, QAP, RCPA HEMATOL, WESTMEAD, NSW, AUSTRALIA
关键词
heparin monitoring; APTT;
D O I
10.1080/00313029600169314
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Full dose heparin therapy is monitored by a variety of laboratory methods, of which the activated partial thromboplastin time (APTT) is the most popular. A large number of APTT reagents are currently available, with different sensitivities to heparin evident in many. Within the literature it is apparent that there is a lack of consensus, and indeed some confusion, regarding the therapeutic ranges for the APTT for standard heparin therapy in the treatment of venous thromboembolic disease. Accordingly we conducted an Australasian survey to evaluate current laboratory and clinical practices in monitoring heparin therapy, to determine the extent of variation in the approach and to stimulate the process of standardisation of acceptable procedures and methodology. Results of the survey demonstrate that currently there is no uniform practice used to establish therapeutic ranges for monitoring standard heparin therapy. Furthermore, results suggest that current practice may lead to subtherapeutic anticoagulation in many laboratories.
引用
收藏
页码:343 / 347
页数:5
相关论文
共 17 条
[1]  
BAIN B, 1980, AM J CLIN PATHOL, V74, P668
[2]   PROSPECTIVE STUDY OF VALUE OF MONITORING HEPARIN TREATMENT WITH ACTIVATED PARTIAL THROMBOPLASTIN TIME [J].
BASU, D ;
CADE, J ;
GALLUS, A ;
HIRSH, J .
NEW ENGLAND JOURNAL OF MEDICINE, 1972, 287 (07) :324-+
[3]  
BRANDT JT, 1981, AM J CLIN PATHOL, V76, P530
[4]   ESTABLISHING A THERAPEUTIC RANGE FOR HEPARIN-THERAPY [J].
BRILLEDWARDS, P ;
GINSBERG, JS ;
JOHNSTON, M ;
HIRSH, J .
ANNALS OF INTERNAL MEDICINE, 1993, 119 (02) :104-109
[5]  
CHIU HM, 1977, BLOOD, V49, P171
[6]  
*COLL AM PATH, 1995, COMPR COAG SURV
[7]  
Favaloro E. J., 1994, Australian Journal of Medical Science, V15, P133
[8]  
GALLUS AS, 1973, LANCET, V2, P806
[9]  
Hirsh J, 1990, Baillieres Clin Haematol, V3, P685, DOI 10.1016/S0950-3536(05)80024-1
[10]   HEPARIN FOR 5 DAYS AS COMPARED WITH 10 DAYS IN THE INITIAL TREATMENT OF PROXIMAL VENOUS THROMBOSIS [J].
HULL, RD ;
RASKOB, GE ;
ROSENBLOOM, D ;
PANJU, AA ;
BRILLEDWARDS, P ;
GINSBERG, JS ;
HIRSH, J ;
MARTIN, GJ ;
GREEN, D .
NEW ENGLAND JOURNAL OF MEDICINE, 1990, 322 (18) :1260-1264