Monte Carlo simulation of a typical 60Co therapy source

被引:165
作者
Mora, GM
Maio, A
Rogers, DWO
机构
[1] Univ Lisbon, CFNUL, P-1649 Lisbon, Portugal
[2] Univ Lisbon, FCUL, P-1649 Lisbon, Portugal
[3] Natl Res Council Canada, Ottawa, ON K1A 0R6, Canada
关键词
D O I
10.1118/1.598770
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The BEAM Monte Carlo code is used to simulate the Co-60 beam from an Eldorado 6 radiotherapy unit and to calculate the relative air-kerma output factors as a function of field size. The unit is realistically modeled, including source capsule, housing and collimator assembly. The calculated relative air-kerma output factors at SSD = 80.5 cm agree to within 0.1% with measured values. It is shown that the variation of the output factor is almost entirely due to scattered photons from the fixed and adjustable collimators and there is no effect of shadowing primary photons. The influence of the geometry of the collimation system on the photon spectra on-axis is shown to be small but finite. The calculated buildup region of a depth-dose curve in a water phantom irradiated by a narrow and a broad Co-60 beam is shown to agree with experimental data at the 2% to 3% level. Unlike previous calculations, the results accurately predict the effects of electron contamination from the surface to dose maximum. The variation of electron contamination with field size is also presented, as are spectra as a function of field size. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)01111-6].
引用
收藏
页码:2494 / 2502
页数:9
相关论文
共 16 条
[1]   SPECTRA OF INTERNALLY SCATTERED RADIATION FROM LARGE CO60 SOURCES USED IN TELETHERAPY [J].
AITKEN, JH ;
HENRY, WH .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1964, 15 (12) :713-&
[2]   ELECTRON CONTAMINATION IN CO-60 GAMMA-RAY BEAMS [J].
ATTIX, FH ;
LOPEZ, F ;
OWOLABI, S ;
PALIWAL, BR .
MEDICAL PHYSICS, 1983, 10 (03) :301-306
[3]  
BIELAJEW AF, 1993, PIRS0396 NAT RES COU
[4]  
BIELAJEW AF, 1989, MONTE CARLO TRANSPOR, P407
[5]   MONTE-CARLO SIMULATION OF A CO-60 BEAM [J].
HAN, K ;
BALLON, D ;
CHUI, C ;
MOHAN, R .
MEDICAL PHYSICS, 1987, 14 (03) :414-419
[6]  
Hubbell JH, 1995, X -Ray Mass Attenuation Coefficients, DOI DOI 10.18434/T4D01F
[7]  
*ICRU, 1971, 18 ICRU
[8]  
KHAN FM, 1980, MED PHYS, V7, P230, DOI 10.1118/1.594648
[9]  
KHAN FM, 1984, PHYSICS RAD THERAPY
[10]  
MA CM, 1995, 509E NRC PIRS