Prospective randomized comparison of cooled radiofrequency versus standard radiofrequency energy for ablation of typical atrial flutter

被引:26
作者
Atiga, WL
Worley, SJ
Hummel, J
Berger, RD
Gohn, DC
Mandalakas, NJ
Kalbfleisch, S
Halperin, H
Donahue, K
Tomaselli, G
Calkins, H
Daoud, E
机构
[1] Lancaster Heart Fdn, Lancaster, PA USA
[2] Univ Pittsburgh, Med Ctr, Pittsburgh, PA 15260 USA
[3] Midwest Cardiol Res Fdn, Columbus, OH USA
[4] Johns Hopkins Univ Hosp, Baltimore, MD 21287 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2002年 / 25卷 / 08期
关键词
cooled; radiofrequency ablation; atrial flutter;
D O I
10.1046/j.1460-9592.2002.01172.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In patients with atrial flutter, conventional RF ablation may not result in complete isthmus block. This prospective, randomized study tested the hypothesis that the cooled RF ablation is safe and facilitates the achievement of isthmus block with fewer RF applications than with standard ablation for typical atria] flutter. Isthmus ablation was performed in 59 patients (40 men, 64 14 years) with type I atrial flutter using standard RF (n = 31) or cooled RF (n 28) catheters with crossover after 12 unsuccessful RF applications. The endpoint was bidirectional isthmus block or a total of 24 unsuccessful RF applications. After the first 12 RF applications, 17 (55%) of 31 standard RF and 22 (79%) of 28 cooled RF patients had bidirectional isthmus block (P < 0.05). After the remaining patients crossed over to the alternate RF ablation system and underwent up to 12 more RF applications, bidirectional isthmus block had been demonstrated in 27 (87%) of 31 standard RF and 25 (89%) of 28 cooled RF patients (P = NS). Isthmus block was not achieved within 24 RF applications in four standard and three cooled RF patients. Mean measured tip temperatures for cooled RF were lower than for standard RF (38.5°C +/- 6.98°C vs 57.2°C +/- 7.42°C, P < 0.0001). Peak temperatures were also lower for cooled RF compared to standard RF (45.7degreesC +/- 22.7degreesC vs 63.4degreesC +/- 9.87degreesC, P < 0.0001). Importantly, mean power delivered was significantly higher for cooled than for standard RF (42.3 +/- 9.48 vs 34.0 +/- 14.0 W, P < 0.0001). There were no serious complications for either ablation system. During a 12.8 +/- 3.76-month follow-up, there were two atrial flutter recurrences in the cooled RF group and four in the standard RF group (P = NS). In patients with type I atrial flutter, ablation with the cooled RF catheter is as safe as, and facilitates creation of bidirectional isthmus block more rapidly than, standard RF ablation.
引用
收藏
页码:1172 / 1178
页数:7
相关论文
共 31 条
[1]   Recurrence of conduction following radiofrequency catheter ablation procedures: Relationship to ablation target and electrode temperature [J].
Calkins, H ;
Prystowsky, E ;
Berger, RD ;
Saul, JP ;
Klein, LS ;
Liem, LB ;
Huang, SKS ;
Gillette, P ;
Yong, P ;
Carlson, M ;
Kay, GN ;
Dailey, S ;
Epstein, A ;
Plumb, V ;
Wharton, JM ;
Kanter, R ;
Sorrentino, R ;
Greenfield, RA ;
Walsh, E ;
Waldo, A ;
Biblo, L ;
Johnson, N ;
Lieberman, R ;
Rosenbaum, D ;
Mackall, J ;
VanHare, G ;
Mittleman, RS ;
Wagshal, AB ;
Greene, T ;
Lauer, M ;
Sung, R ;
Young, C ;
Evans, J ;
Hackett, K ;
Miles, W ;
Zipes, D ;
Swartz, J ;
Lawrence, J ;
Tomaselli, G .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (08) :704-712
[2]   CATHETER ABLATION OF ATRIAL-FLUTTER USING RADIOFREQUENCY ENERGY [J].
CALKINS, H ;
LEON, AR ;
DEAM, AG ;
KALBFLEISCH, SJ ;
LANGBERG, JJ ;
MORADY, F .
AMERICAN JOURNAL OF CARDIOLOGY, 1994, 73 (05) :353-356
[3]   Electrophysiological effects of catheter ablation of inferior vena cava-tricuspid annulus isthmus in common atrial flutter [J].
Cauchemez, B ;
Haissaguerre, M ;
Fischer, B ;
Thomas, O ;
Clementy, J ;
Coumel, P .
CIRCULATION, 1996, 93 (02) :284-294
[4]  
CHAUVIN M, 1998, PACE 2, V21, P172
[5]   RADIOFREQUENCY ABLATION OF THE INFERIOR VENA-CAVA - TRICUSPID-VALVE ISTHMUS IN COMMON ATRIAL-FLUTTER [J].
COSIO, FG ;
LOPEZGIL, M ;
GOICOLEA, A ;
ARRIBAS, F ;
BARROSO, JL .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (08) :705-709
[6]   Ablation of ventricular tachycardia with a saline-cooled radiofrequency catheter: Anatomic and histologic characteristics of the lesions in humans [J].
Delacretaz, E ;
Stevenson, WG ;
Winters, GL ;
Mitchell, RN ;
Stewart, S ;
Lynch, K ;
Friedman, PL .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1999, 10 (06) :860-865
[7]   RADIOFREQUENCY CATHETER ABLATION FOR THE TREATMENT OF HUMAN TYPE-1 ATRIAL-FLUTTER - IDENTIFICATION OF A CRITICAL ZONE IN THE REENTRANT CIRCUIT BY ENDOCARDIAL MAPPING TECHNIQUES [J].
FELD, GK ;
FLECK, RP ;
CHEN, PS ;
BOYCE, K ;
BAHNSON, TD ;
STEIN, JB ;
CALISI, CM ;
IBARRA, M .
CIRCULATION, 1992, 86 (04) :1233-1240
[8]   Radiofrequency catheter ablation of common atrial flutter in 200 patients [J].
Fischer, B ;
Jais, P ;
Shah, DP ;
Chouairi, S ;
Haissaguerre, M ;
Garrigues, S ;
Poquet, F ;
Gencel, L ;
Clementy, J ;
Marcus, FI .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (12) :1225-1233
[9]  
FISHER B, 1995, J AM COLL CARDIOL, V25, P1365
[10]   OBSERVATIONS ON ELECTRODE-TISSUE INTERFACE TEMPERATURE AND EFFECT ON ELECTRICAL-IMPEDANCE DURING RADIOFREQUENCY ABLATION OF VENTRICULAR MYOCARDIUM [J].
HAINES, DE ;
VEROW, AF .
CIRCULATION, 1990, 82 (03) :1034-1038