Erroneous intracranial pressure measurements from simultaneous pressure monitoring and ventricular drainage catheters

被引:25
作者
Birch, A. A.
Eynon, C. A.
Schley, D.
机构
[1] Southampton Gen Hosp, Dept Med Phys & Bioengn, Southampton SO16 6YD, Hants, England
[2] Southampton Gen Hosp, Wessex Neurol Ctr, Southampton SO16 6YD, Hants, England
[3] Univ Jaffna, Dept Math & Stat, Jaffna, Sri Lanka
关键词
intracranial pressure; cerebrospinal fluid; ventricular drainage;
D O I
10.1385/NCC:5:1:51
中图分类号
R4 [临床医学];
学科分类号
1002 [临床医学]; 100602 [中西医结合临床];
摘要
The objective of this report is to highlight the potential for false pressure measurements from systems that combine intracranial pressure (ICP) measurement and ventricular drainage. If the ports of the drain become blocked to the extent that they present a high resistance to cerebrospinal fluid flow, then a significant pressure gradient between the inside and outside of the catheter may be established. Thus, any intracatheter transducer will faithfully record a pressure much lower than true ICP. This holds true for catheter-tip transducers when the transducer lies inside the catheter. In the absence of flow, however, pressures will equalize; therefore, accurate measurements may be taken if the drain is temporarily closed. We model this situation and provide simulations of expected measurements in such situations; these compare well to observed clinical readings.
引用
收藏
页码:51 / 54
页数:4
相关论文
共 5 条
[1]
Knowlton FP, 1912, J PHYSIOL-LONDON, V44, P206
[2]
A STARLING RESISTOR REGULATES CEREBRAL VENOUS OUTFLOW IN DOGS [J].
LUCE, JM ;
HUSEBY, JS ;
KIRK, W ;
BUTLER, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1982, 53 (06) :1496-1503
[3]
Pressure-controlled drainage of cerebrospinal fluid: Clinical experience with a new type of ventricular catheter (Ventcontrol MTC) and an integrated Piezo-resistive sensor at its tip: Technical note [J].
Piek, J ;
Raes, P .
NEUROSURGERY, 1996, 38 (01) :216-218
[4]
A model of in-vivo hydrocephalus shunt dynamics for blockage and performance diagnostics [J].
Schley, D ;
Billingham, J ;
Marchbanks, RJ .
MATHEMATICAL MEDICINE AND BIOLOGY-A JOURNAL OF THE IMA, 2004, 21 (04) :347-368
[5]
ERRONEOUS MEASUREMENT OF INTRACRANIAL-PRESSURE CAUSED BY SIMULTANEOUS VENTRICULAR DRAINAGE - A HYDRODYNAMIC MODEL STUDY [J].
WILKINSON, HA ;
YARZEBSKI, J ;
WILKINSON, EC ;
ANDERSON, FA .
NEUROSURGERY, 1989, 24 (03) :348-354