An investigation of the physical characteristics of Ga-66 as an isotope for PET imaging and quantification

被引:34
作者
Graham, MC [1 ]
Pentlow, KS [1 ]
Mawlawi, O [1 ]
Finn, RD [1 ]
Daghighian, F [1 ]
Larson, SM [1 ]
机构
[1] MEM SLOAN KETTERING CANC CTR,DEPT RADIOL,NEW YORK,NY 10021
关键词
gallium-66; Ga-66; PET quantification; monoclonal antibody dosimetry; positron emission tomography (PET);
D O I
10.1118/1.597924
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Isotopes commonly used for PET imaging and quantification have a straightforward decay scheme involving ''pure'' positron (beta(+)) emission, i.e., 95%-100% beta(+) abundance, with no additional gamma rays. Ga-66 (E(max)=4.2 MeV, T-1/2=9.5 h) is a member of a category of isotopes with a lower abundance of beta(+,)s (57%) and a more complicated spectrum involving combinations of gamma rays that are emitted in cascade. These additional gamma rays tend to cause a higher singles rate, resulting in more random coincidence events. The most abundant positron (51.5%) in the spectrum has one of the highest energies considered for PET imaging. For the purposes of monoclonal antibody dosimetry using Ga-66, it is important to verify the quantification in phantoms prior to initiating human studies. A series of quantitative phantom measurements were performed on the PC4600, a head-optimized EGO based scanner with multiple detector rings. Count rate linearity was verified over concentrations ranging from 4.0 kBq/cc to 37 kBq/cc (0.11-1.0 mu Ci/cc); resolution averaged 16 mm full width half-maximum in the x and y directions in both the direct and cross planes. Axial resolution was 14 mm. The range of the energetic positrons (up to 4.153 MeV, range 7.6 mm in tissue) was verified as a primary source of resolution degradation. Within the limits outlined above, Ga-66 is a suitable isotope for use as Ga-66 citrate or with monoclonal antibodies in the detection and staging of tumors and other lesions. In addition, the energetic positrons have possible therapeutic applications when used as a monoclonal antibody label. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:317 / 326
页数:10
相关论文
共 40 条
[1]  
[Anonymous], PRECISION MEASUREMEN
[2]   CORRECTIONS FOR ATTENUATION, SCATTERED RADIATION, AND RANDOM COINCIDENCES IN A RING DETECTOR POSITRON EMISSION TRANSAXIAL TOMOGRAPH [J].
BERGSTROM, M ;
BOHM, C ;
ERICSON, K ;
ERIKSSON, L ;
LITTON, J .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1980, 27 (01) :549-554
[3]  
BIGLER RE, 1984, MED PHYSICS MONOGRAP, V10, P411
[4]  
BROWNE E, 1986, TABEL RADIOACTIVE IS
[5]  
CARROLL L, COMMUNICATION
[6]  
Carroll LR, 1983, EMISSION COMPUTED TO, P235
[7]  
CHO ZH, 1975, J NUCL MED, V16, P1174
[8]  
DAGHIGHIAN F, 1993, EUR J NUCL MED, V20, P402
[9]   A POSITRON TOMOGRAPH WITH 600-BGO CRYSTALS AND 2.6 MM RESOLUTION [J].
DERENZO, SE ;
HUESMAN, RH ;
CAHOON, JL ;
GEYER, AB ;
MOSES, WW ;
UBER, DC ;
VULETICH, T ;
BUDINGER, TF .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :659-664
[10]  
DILLMAN LT, 1975, 10 MIRD SOC NUCL MED