An improved model for the energy resolution of multicrystal encoding detectors for PET

被引:8
作者
Vozza, D
Moisan, C
Paquet, S
机构
[1] TRIUMF, Vancouver
关键词
modeling; nonproportionality; PET detectors; scintillator;
D O I
10.1109/23.568800
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Recently, we have shown that the program DETECT provides a realistic platform to model the performances of scintillation counters with geometries as involved as that of the EXACT HR PLUS, a state-of-the-art position and energy encoding EGO detector manufactured by Siemens-CTI for positron emission tomography (PET), In this paper we present the results of new simulations of this block's performances, updated to include the nonproportionality of the scintillation response of EGO and the statistical noise from photoelectron amplification in photomultiplier tubes (PMT's), The updated model accounts for the crystal-dependent energy resolutions observed with 511-keV photons in the detector with excellent accuracy, Measured and simulated resolutions are shown to agree within -1 to 4%, with an average absolute difference of 1% over the entire block. The impact of the nonproportional scintillation response of EGO on the energy resolution is shown to be relatively small compared to that of the amplification noise in the four PMT's reading out the block.
引用
收藏
页码:179 / 183
页数:5
相关论文
共 25 条
[1]   FLUORESCENT RESPONSE OF NAI(T1) CSI(T1) CSI(NA) AND CAF2(EU) TO X-RAYS AND LOW ENERGY GAMMA RAYS [J].
AITKEN, DW ;
BERON, BL ;
YENICAY, G ;
ZULLIGER, HR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1967, NS14 (01) :468-+
[2]  
[Anonymous], PRIB TEKH EKSP
[3]  
[Anonymous], 1964, THEORY PRACTICE SCIN
[4]   MONTE-CARLO SIMULATION OF BAF2 DETECTORS USED IN TIME-OF-FLIGHT POSITRON EMISSION TOMOGRAPHY [J].
BICE, AN ;
LEWELLEN, TK ;
MIYAOKA, RS ;
HARRISON, RL ;
HAYNOR, DR ;
POLLARD, KR ;
ANSON, CP ;
GILLISPIE, SB .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :696-701
[5]  
*BURLC IND INC, 1980, PHOT HDB
[6]   MONTE-CARLO OPTIMIZATION OF DEPTH-OF-INTERACTION RESOLUTION IN PET CRYSTALS [J].
DEVOL, TA ;
MOSES, WW ;
DERENZO, SE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (02) :170-174
[7]   Non-proportionality in the scintillation response and the energy resolution obtainable with scintillation crystals [J].
Dorenbos, P ;
deHaas, JTM ;
vanEijk, CWE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (06) :2190-2202
[8]   A MEASUREMENT OF THE LIGHT YIELD OF COMMON INORGANIC SCINTILLATORS [J].
HOLL, I ;
LORENZ, E ;
MAGERAS, G .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :105-109
[9]   LIGHT COLLECTION IN SCINTILLATION DETECTOR COMPOSITES FOR NEUTRON DETECTION [J].
KNOLL, GF ;
KNOLL, TF ;
HENDERSON, TM .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :872-875
[10]  
MCDANIEL D, 1993, P 1993 IEEE NUCL SCI, P1739