Injection pressures by anesthesiologists during simulated peripheral nerve block

被引:67
作者
Claudio, R [1 ]
Hadzic, A [1 ]
Shih, H [1 ]
Vloka, JD [1 ]
Castro, J [1 ]
Koscielniak-Nielsen, Z [1 ]
Thys, DM [1 ]
Santos, AC [1 ]
机构
[1] Columbia Univ, St Lukes Roosevelt Hosp Ctr, Dept Anesthesiol, Coll Phys & Surg, New York, NY 10025 USA
关键词
regional anesthesia; peripheral nerve block; neurologic complications; needle;
D O I
10.1016/j.rapm.2003.12.013
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background and Objectives: Anesthesiologists typically rely on a subjective evaluation ("syringe feel") of possible abnormal resistance to injection while performing a peripheral nerve block (PNB). A greater force required to perform the injection is believed to be associated with intraneural injection. The hypothesis of this study is that anesthesiologists vary in their perception of "normal" injection force, that the syringe feel method is inconsistent in estimating resistance, and that needle design may affect the injection force. Methods: Thirty anesthesiologists were asked to inject a local anesthetic, as they would in their everyday practice, through a commonly used syringe and needle assembly. Injection force was measured using an in-line manometer coupled to a computer via an analog-to-digital conversion board. In addition, injection force at clinically relevant injection speeds was determined using 3 differently sized needles from 4 different manufacturers. Results: During a steady injection rate, all anesthesiologists perceived an increase in the force required to inject, even with minor pressures changes (0.6 +/- 0.3 psi). However, the 30 anesthesiologists, 21 (70%) initiated injection using a force that resulted in pressures greater than 20 psi; 15 (50%) used a force greater than 25 psi, and 3 (10%) exerted pressures greater than 30 psi. Pressures varied as much as 20-fold among needles of the same gauge/length from different manufacturers (P < .01). Conclusions: Anesthesiologists vary widely in their perception of appropriate force and rate of injection during PNB. The syringe-feel method of assessing injection force is inconsistent and may be further affected by variability in needle design.
引用
收藏
页码:201 / 205
页数:5
相关论文
共 13 条
[1]  
Adriani J, 1985, LABATS REGIONAL ANES, P107
[2]   Major complications of regional anesthesia in France - The SOS Regional Anesthesia Hotline Service [J].
Auroy, Y ;
Benhamou, D ;
Bargues, L ;
Ecoffey, C ;
Falissard, B ;
Mercier, F ;
Bouaziz, H ;
Samii, K .
ANESTHESIOLOGY, 2002, 97 (05) :1274-1280
[3]   Central nervous system toxicity following the administration of levobupivacaine for lumbar plexus block: A report of two cases [J].
Breslin, DS ;
Martin, G ;
MacLeod, DB ;
D'Ercole, F ;
Grant, SA .
REGIONAL ANESTHESIA AND PAIN MEDICINE, 2003, 28 (02) :144-147
[4]  
Crews JC, 2003, ANESTH ANALG, V96, P1188, DOI [10.1213/01.ane.0000053235.07186.39, 10.1213/01.ANE.0000053235.07186.39]
[5]  
HADZIC A, 2003, IMRAPT, V15, P3
[6]  
JANKOVIC D, 2001, BRACHIAL PLEXUS REGI, P76
[7]   Debilitating chronic pain syndromes after presumed intraneural injections [J].
Kaufman, BR ;
Nystrom, E ;
Nath, S ;
Foucher, G ;
Nystrom, A .
PAIN, 2000, 85 (1-2) :283-286
[8]   Successful resuscitation after ropivacaine-induced ventricular fibrillation [J].
Klein, SM ;
Pierce, T ;
Rubin, Y ;
Nielsen, KC ;
Steele, SM .
ANESTHESIA AND ANALGESIA, 2003, 97 (03) :901-903
[9]   Convulsions induced by ropivacaine during interscalene brachial plexus block [J].
Korman, B ;
Riley, RH .
ANESTHESIA AND ANALGESIA, 1997, 85 (05) :1128-1129
[10]   LONGITUDINAL SPREAD OF INTRANEURALLY INJECTED LOCAL-ANESTHETICS - EXPERIMENTAL-STUDY OF INITIAL NEURAL DISTRIBUTION FOLLOWING INTRANEURAL INJECTIONS [J].
SELANDER, D ;
SJOSTRAND, J .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1978, 22 (06) :622-634