Ceramic-ceramic and metal-polyethylene total hip replacements - Comparison of pseudomembranes after loosening

被引:95
作者
Lerouge, S
Huk, O
Yahia, L
Witvoet, J
Sedel, L
机构
[1] ECOLE POLYTECH, INST BIOMED ENGN, GRP RES BIOMECH & BIOMAT, MONTREAL, PQ H3C 3A7, CANADA
[2] HOP ST LOUIS, PARIS, FRANCE
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1997年 / 79B卷 / 01期
关键词
D O I
10.1302/0301-620X.79B1.6621
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We made a semiquantitative study of the comparative histology of pseudomembranes from 12 loose cemented ceramic-ceramic and 18 metal-polyethylene total hip replacements, We found no significant difference in cellular reaction between the two groups, but there was a major difference ih the origin of the particulate debris, In the metal-polyethylene group, polyethylene of articular origin was predominant, while in the ceramic-ceramic group the cellular reaction appeared to be a response to zirconia ceramic particles used to opacify cement used for fixation. Isolation and characterisation of the debris showed that the zirconia particles formed the greatest proportion (76%) in ceramic-ceramic hips, while alumina debris of articular origin formed only 12%. Our study has indicated that aseptic loosening of ceramic cups is not due to a response to debris generated at the articular interface, but to mechanical factors which lead to fragmentation of the cement.
引用
收藏
页码:135 / 139
页数:5
相关论文
共 16 条
[1]   METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]  
BOUTIN P, 1972, REV CHIR ORTHOP, V58, P229
[3]   THE USE OF DENSE ALUMINA ALUMINA CERAMIC COMBINATION IN TOTAL HIP-REPLACEMENT [J].
BOUTIN, P ;
CHRISTEL, P ;
DORLOT, JM ;
MEUNIER, A ;
DEROQUANCOURT, A ;
BLANQUAERT, D ;
HERMAN, S ;
SEDEL, L ;
WITVOET, J .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1988, 22 (12) :1203-1232
[4]  
CROLET JM, 1989, THESIS U TECHNOLOGIE
[5]   WEAR ANALYSIS OF RETRIEVED ALUMINA HEADS AND SOCKETS OF HIP PROSTHESES [J].
DORLOT, JM ;
CHRISTEL, P ;
MEUNIER, A .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH-APPLIED BIOMATERIALS, 1989, 23 (A3) :299-310
[6]  
GLANT TT, 1993, J BONE MINER RES, V8, P1071
[7]  
GOODMAN SB, 1992, AM SOC TEST MATER, V1144, P27, DOI 10.1520/STP14849S
[8]   POLYETHYLENE AND METAL DEBRIS GENERATED BY NON-ARTICULATING SURFACES OF MODULAR ACETABULAR COMPONENTS [J].
HUK, OL ;
MANJULA, B ;
BETTS, F ;
RIMNAC, CM ;
LIEBERMAN, JR ;
HUO, MH ;
SALVATI, EA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (04) :568-574
[9]  
LEROUGE S, 1996, IN PRESS J BIOMED MA
[10]   ISOLATION AND CHARACTERIZATION OF WEAR PARTICLES GENERATED IN PATIENTS WHO HAVE HAD FAILURE OF A HIP-ARTHROPLASTY WITHOUT CEMENT [J].
MALONEY, WJ ;
SMITH, RL ;
SCHMALZRIED, TP ;
CHIBA, J ;
HUENE, D ;
RUBASH, H .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77 (09) :1301-1310