Investigation of chloride sensitive ISFETs with different membrane compositions suitable for medical applications

被引:52
作者
Bratov, A [1 ]
Abramova, N [1 ]
Domínguez, C [1 ]
机构
[1] UAB, CNM, Grupo Transductores Quim, E-08193 Bellaterra, Spain
关键词
ISFET; photocured polymer membrane; Cl-ion sensor; clinical analysis;
D O I
10.1016/j.aca.2004.01.034
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Cl--ion sensitive ISFETs with photocured polyurethane-based polymer membranes with six different ionophores (ETH 9033, ETH 9009, MnTPPCl, organotin compounds and traditional ion-exchanger TDMACl) have been studied in pure NaCl solutions and in background solutions containing anions in concentrations normally found in a whole blood and serum. PVC or silicon resin were used to form the membranes in cases when it was not possible to use the photocurable polymer composition. Experimental results on determination of chloride ions in serum samples are presented. Performed tests showed that all chloride-selective ionophores commercially proposed up to now, both neutral and charged carriers, do not provide better selectivity and stability of chloride ion sensors for clinical application than the traditional anionic ion-exchanger TDMACl. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:99 / 106
页数:8
相关论文
共 35 条
[1]   Application of an ion-selective field effect transistor with a photocured polymer membrane in nephrology for determination of potassium ions in dialysis solutions and in blood plasma [J].
Abramova, N ;
Borisov, Y ;
Bratov, A ;
Gavrilenko, P ;
Domínguez, C ;
Spiridonov, V ;
Suglobova, E .
TALANTA, 2000, 52 (03) :533-538
[2]   Cationic or anionic sites?: Selectivity optimization of ion-selective electrodes based on charged ionophores [J].
Amemiya, S ;
Bühlmann, P ;
Pretsch, E ;
Rusterholz, B ;
Umezawa, Y .
ANALYTICAL CHEMISTRY, 2000, 72 (07) :1618-1631
[3]   POTENTIOMETRIC ANION SELECTIVITY OF POLYMER MEMBRANES DOPED WITH PALLADIUM ORGANOPHOSPHINE COMPLEX [J].
BADR, IHA ;
MEYERHOFF, ME ;
HASSAN, SSM .
ANALYTICAL CHEMISTRY, 1995, 67 (15) :2613-2618
[4]   ANION-SELECTIVE MEMBRANE ELECTRODES BASED ON METALLOPORPHYRINS - THE INFLUENCE OF LIPOPHILIC ANIONIC AND CATIONIC SITES ON POTENTIOMETRIC SELECTIVITY [J].
BAKKER, E ;
MALINOWSKA, E ;
SCHILLER, RD ;
MEYERHOFF, ME .
TALANTA, 1994, 41 (06) :881-890
[5]  
Baldi A, 1999, QUIM ANAL, V18, P43
[6]   Optimization of photocurable polyurethane membrane composition for ammonium ion sensor [J].
Bratov, A ;
Abramova, N ;
Munoz, J ;
Dominguez, C ;
Alegret, S ;
Batroli, J .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1997, 144 (02) :617-621
[7]   Ion-selective field effect transistor (ISFET)-based calcium ion sensor with photocured polyurethane membrane suitable for ionised calcium determination in milk [J].
Bratov, A ;
Abramova, N ;
Domínguez, C ;
Baldi, A .
ANALYTICA CHIMICA ACTA, 2000, 408 (1-2) :57-64
[8]   PHOTOCURABLE POLYMER MATRICES FOR POTASSIUM-SENSITIVE ION-SELECTIVE ELECTRODE MEMBRANES [J].
BRATOV, A ;
ABRAMOVA, N ;
MUNOZ, J ;
DOMINGUEZ, C ;
ALEGRET, S ;
BARTROLI, J .
ANALYTICAL CHEMISTRY, 1995, 67 (19) :3589-3595
[9]   Carrier-based ion-selective electrodes and bulk optodes. 2. Ionophores for potentiometric and optical sensors [J].
Buhlmann, P ;
Pretsch, E ;
Bakker, E .
CHEMICAL REVIEWS, 1998, 98 (04) :1593-1687
[10]   ASYMMETRIC CELLULOSE-ACETATE MEMBRANE-BASED CARBONATE-SELECTIVE AND CHLORIDE-SELECTIVE ELECTRODES [J].
CHA, MJ ;
SHIN, JH ;
OH, BK ;
KIM, CY ;
CHA, GS ;
SHIN, DS ;
KIM, B .
ANALYTICA CHIMICA ACTA, 1995, 315 (03) :311-319