Surface plasmon resonance analysis of aqueous mercuric ions

被引:79
作者
Chah, S
Yi, JH
Zare, RN
机构
[1] Stanford Univ, Dept Chem, Stanford, CA 94305 USA
[2] Seoul Natl Univ, Sch Chem Engn, Seoul 151744, South Korea
关键词
mercuric ions; SPR; 1,6-hexanedithiol; selective detection;
D O I
10.1016/j.snb.2003.11.015
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Mercuric ions, Hg[II], can be detected and quantitated by measuring the surface plasmon resonance (SPR) signal when a gold film treated with 1,6-hexanedithiol (HDT), which produced a self-assembled monolayer, is exposed to mercury ions in aqueous solution. Changes in the SPR signal allowed not only quantification of the concentration of Hg[II] in solution from 1.0 nM to 1.0 mM but also selective detection of Hg[II] ions when present in mixtures containing Pb[II], Ni[II], Zn[II], and Cu[II] ions. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:216 / 222
页数:7
相关论文
共 22 条
[1]   Surface plasmon resonance fibre-optic sensor for gas detection [J].
Abdelghani, A ;
Chovelon, JM ;
JaffrezicRenault, N ;
RonotTrioli, C ;
Veillas, C ;
Gagnaire, H .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 39 (1-3) :407-410
[2]   Fluorometric chemosensors.: Interaction of toxic heavy metal ions PbII, CdII, and HgII with novel mixed-donor phenanthroline-containing macrocycles:: Spectrofluorometric, conductometric, and crystallographic studies [J].
Aragoni, MC ;
Arca, M ;
Demartin, F ;
Devillanova, FA ;
Isaia, F ;
Garau, A ;
Lippolis, V ;
Jalali, F ;
Papke, U ;
Shamsipur, M ;
Tei, L ;
Yari, A ;
Verani, G .
INORGANIC CHEMISTRY, 2002, 41 (25) :6623-6632
[3]  
ARAKAWA T, 1996, THIN SOLID FILMS, V281, P466
[4]   High-resolution multiwavelength surface plasmon resonance spectroscopy for probing conformational and electronic changes in redox proteins [J].
Boussaad, S ;
Pean, J ;
Tao, NJ .
ANALYTICAL CHEMISTRY, 2000, 72 (01) :222-226
[5]   Separation of zinc ions from aqueous solutions using modified silica impregnated with CYANEX 272 [J].
Chah, S ;
Kim, JS ;
Yi, JH .
SEPARATION SCIENCE AND TECHNOLOGY, 2002, 37 (03) :701-716
[6]  
Chah S, 2002, LANGMUIR, V18, P314, DOI 10.1021/1011226y
[7]   Integrated analytical sensors: the use of the TISPR-1 as a biosensor [J].
Elkind, JL ;
Stimpson, DI ;
Strong, AA ;
Bartholomew, DU ;
Melendez, JL .
SENSORS AND ACTUATORS B-CHEMICAL, 1999, 54 (1-2) :182-190
[8]   Studies of adsorption kinetics and defects of self-assembled thiol monolayers on gold by capacitance plane plot [J].
Guo, M ;
Diao, P ;
Tong, P .
JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 2000, 47 (06) :1197-1203
[9]   Colloidal Au-enhanced surface plasmon resonance for ultrasensitive detection of DNA hybridization [J].
He, L ;
Musick, MD ;
Nicewarner, SR ;
Salinas, FG ;
Benkovic, SJ ;
Natan, MJ ;
Keating, CD .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2000, 122 (38) :9071-9077
[10]   Mercury chloride modulation of the GABA(A) receptor-channel complex in rat dorsal root ganglion neurons [J].
Huang, CS ;
Narahashi, T .
TOXICOLOGY AND APPLIED PHARMACOLOGY, 1996, 140 (02) :508-520