Numerical simulation of nonpulsatile left ventricular bypass

被引:4
作者
Mitsui, N
Fukunaga, S
Koura, Y
Morita, S
Hotei, H
Sueshiro, M
Sueda, T
Matsuura, Y
机构
[1] First Department of Surgery, Hiroshima University, School of Medicine, Hiroshima
[2] First Department of Surgery, Hiroshima University, School of Medicine, Minami-ku, Hiroshima 734
关键词
computer simulation; nonpulsatile blood pump; hemodynamics; heart failure; variable compliance; left ventricular bypass;
D O I
10.1111/j.1525-1594.1996.tb04494.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
A computer simulation was carried out to investigate the influence of nonpulsatile left ventricular assistance on hemodynamics. A simulation circuit was constructed to represent the circulatory system. A source of current was added to denote the nonpulsatile blood pump. The left and right ventricles were replaced by variable compliances. Left heart failure was simulated by decreasing the amount of compliance change of the left ventricle, We introduced a pulsatility indicator (PI) to clarify the pulsatility characteristics in the hemodynamics; this PI was defined as the ratio of the pulse pressure (PP) to the mean aortic pressure (AoP). When nonpulsatile bypass flow increased, the mean AoP, tension time index (TTI), and diastolic pressure time index (DPTI) increased, and cardiac output, PP, and PI decreased. When assisted flow increased with the constant total flow rate, the mean AoP and DPTI changed little; the PP, TTI, and PI decreased, and the endocardial viability rate increased. The PI would be helpful in evaluating the effect of pulsatility.
引用
收藏
页码:632 / 636
页数:5
相关论文
共 14 条
[1]
EXPERIMENTAL SUBENDOCARDIAL ISCHEMIA IN DOGS WITH NORMAL CORONARY-ARTERIES [J].
BUCKBERG, GD ;
ARCHIE, JP ;
FIXLER, DE ;
HOFFMAN, JIE .
CIRCULATION RESEARCH, 1972, 30 (01) :67-+
[2]
A MODULAR NUMERICAL-MODEL OF THE CARDIOVASCULAR-SYSTEM FOR STUDYING AND TRAINING IN THE FIELD OF CARDIOVASCULAR PHYSIOPATHOLOGY [J].
FERRARI, G ;
DELAZZARI, C ;
MIMMO, R ;
TOSTI, G ;
AMBROSI, D .
JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (02) :91-107
[3]
FUKUNAGA S, 1992, TECHN M LIN DRI IEEJ, P1
[4]
HAYASHI S, 1991, HIROSHIMA MED J, V39, P15
[5]
MARUYAMA T, 1993, HIROSHIMA MED J, V41, P143
[6]
MIZOGUCHI K, 1995, ARTIF ORGANS, V19, P653
[7]
A SIMULATION SUPPORT SYSTEM FOR SOLVING LARGE PHYSIOLOGICAL MODELS ON MICROCOMPUTERS [J].
MONTANI, JP ;
ADAIR, TH ;
SUMMERS, RL ;
COLEMAN, TG ;
GUYTON, AC .
INTERNATIONAL JOURNAL OF BIO-MEDICAL COMPUTING, 1989, 24 (01) :41-54
[8]
PAE WE, 1992, J THORAC CARDIOV SUR, V104, P541
[9]
PHILIPS PA, 1975, J THORAC CARDIOV SUR, V69, P30
[10]
Qian K X, 1994, ASAIO J, V40, P213