Thresholds for activation of rabbit retinal ganglion cells with a subretinal electrode

被引:92
作者
Jensen, Ralph J.
Rizzo, Joseph F., III
机构
[1] VA Boston Healthcare Syst, Ctr Innovat Visual Rehabil, Boston, MA 02130 USA
[2] Harvard Univ, Sch Med, Massachusetts Eye & Ear Infirm, Boston, MA 02114 USA
关键词
retinal ganglion cell; electrophysiology; electrical stimulation; retinal prosthesis;
D O I
10.1016/j.exer.2006.01.012
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The ultimate success of a retinal prosthesis to create vision will likely depend upon developing a base of knowledge of how best to electrically stimulate the retina. Previously, we studied the responses of rabbit retinal ganglion cells (RGCs) to current pulses applied with an electrode placed on the epiretinal surface. In the present study, we examined the responses of rabbit RGCs to current pulses applied with a subretinal electrode. Single-unit extracellular recordings were made from OFF RGCs and ON RGCs in isolated retinas, which were stimulated with monophasic current pulses (0.1-50 ms in duration), delivered through a 500-mu m diameter electrode. All RGCs elicited one or more bursts of action potentials upon electrical stimulation of the retina. The timing of the bursts depended upon both the polarity of the electrical stimulus and the RGC type. With near-threshold current pulses, the response latencies of OFF RGCs to anodal stimulation were comparable to those of ON RGCs to cathodal stimulation, whereas the response latencies of OFF RGCs to cathodal stimulation were comparable to those of ON RGCs to anodal stimulation. Threshold currents for activation of RGCs decreased with increased pulse duration. For OFF RGCs, threshold currents for cathodal current pulses were, on average, 2-7.5 times higher (depending upon pulse duration) than the threshold currents for anodal current pulses. For ON RGCs, threshold currents for cathodal and anodal current pulses were, on average, nearly identical for all pulse durations and were equivalent to threshold currents of OFF RGCs to anodal stimulation. With respect to a subretinal prosthesis, our findings suggest the possibility that cathodal current pulses may bias activation of ON RGCs in blind patients. Published by Elsevier Ltd.
引用
收藏
页码:367 / 373
页数:7
相关论文
共 16 条
[1]   NOXIOUS EFFECTS OF EXCESSIVE CURRENTS USED FOR INTRACORTICAL MICROSTIMULATION [J].
ASANUMA, H ;
ARNOLD, AP .
BRAIN RESEARCH, 1975, 96 (01) :103-107
[2]   Artificial human vision [J].
Dowling, J .
EXPERT REVIEW OF MEDICAL DEVICES, 2005, 2 (01) :73-85
[3]   Responses of rabbit retinal ganglion cells to electrical stimulation with an epiretinal electrode [J].
Jensen, Ralph J. ;
Ziv, Ofer R. ;
Rizzo, Joseph F. .
JOURNAL OF NEURAL ENGINEERING, 2005, 2 (01) :S16-S21
[4]   Thresholds for activation of rabbit retinal ganglion cells with relatively large, extracellular microelectrodes [J].
Jensen, RJ ;
Ziv, OR ;
Rizzo, JF .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (04) :1486-1496
[5]   Thresholds for activation of rabbit retinal, ganglion cells with an ultrafine, extracellular microelectrode [J].
Jensen, RJ ;
Rizzo, JF ;
Ziv, OR ;
Grumet, A ;
Wyatt, J .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (08) :3533-3543
[6]  
JENSEN RJ, 1989, VISUAL NEUROSCI, V3, P573
[7]   Electrophysiological studies of the feasibility of suprachoroidal-transretinal stimulation for artificial vision in normal and RCS rats [J].
Kanda, H ;
Morimoto, T ;
Fujikado, T ;
Tano, Y ;
Fukuda, Y ;
Sawai, H .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2004, 45 (02) :560-566
[8]   Selective microstimulation of central nervous system neurons [J].
McIntyre, CC ;
Grill, WM .
ANNALS OF BIOMEDICAL ENGINEERING, 2000, 28 (03) :219-233
[9]   Electrical stimulation of excitable tissue: design of efficacious and safe protocols [J].
Merrill, DR ;
Bikson, M ;
Jefferys, JGR .
JOURNAL OF NEUROSCIENCE METHODS, 2005, 141 (02) :171-198
[10]  
Prince JH, 1964, RABBIT EYE RES