Validation of a solid phase extraction-high performance liquid chromatographic method for the determination of eprosartan in human plasma

被引:14
作者
Ferreiros, N. [1 ]
Iriarte, G. [1 ]
Alonso, R. M. [1 ]
Jimenez, R. M. [1 ]
Ortiz, E. [1 ]
机构
[1] Univ Basque Country, Dept Especialidades Med Quirurgicas, Fac Med & Odontol, EHU, E-48080 Bilbao, Spain
关键词
eprosartan; validation; human plasma; liquid chromatography; solid phase extraction;
D O I
10.1016/j.chroma.2006.02.055
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In this work, a solid phase extraction-reversed phase high performance liquid chromatographic (SPE-PP-HPLC) method with photometric detection for monitoring the antihypertensive drug eprosartan has been validated in order to assure good quantitation of eprosartan in plasma samples obtained from patients under cardiovascular treatment. This analytical method was developed by using experimental design and quantitation was accomplished with the internal standard method. No interferences were observed from endoggenous compounds of plasma and other drugs which are commonly co-administered in elderly patients. The recoveries of eprosartan from plasma samples, measured at three levels of the linear concentration range (150-4000 ng/mL) were found to be between 93.4 and 102.8%. The intraday and interday precision and accuracy (measured by relative standard deviation, RSD, and relative error, RE, respectively) were always lower than 13% (RSD) and 4% (RE). Stability studies showed that eprosartan stock solutions are stable for at least 3 months when stored at 8 degrees C and plasma samples containing the drug were stable at least during the whole analytical method. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:309 / 314
页数:6
相关论文
共 19 条
[1]  
[Anonymous], GUID IND BIOAN METH
[2]  
[Anonymous], 1994, INT C HARM TECHN REQ
[3]  
[Anonymous], 1996, INT C HARM TECHN REQ
[4]   A strategy for validation of bioanalytical methods [J].
Braggio, S ;
Barnaby, RJ ;
Grossi, P ;
Cugola, M .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1996, 14 (04) :375-388
[5]   Experimental design methodologies to optimise the spectrofluorimetric determination of Losartan and Valsartan in human urine [J].
Cagigal, E ;
González, L ;
Alonso, RM ;
Jiménez, RM .
TALANTA, 2001, 54 (06) :1121-1133
[6]  
CHAMBERLAIN J, 1995, SPECIAL PROBLEMS BIO, P35
[7]  
FERREIROS N, IN PRESS
[8]   Fast screening method for the determination of angiotensin II receptor antagonists in human plasma by high-performance liquid chromatography with fluorimetric detection [J].
González, L ;
López, JA ;
Alonso, RM ;
Jiménez, RM .
JOURNAL OF CHROMATOGRAPHY A, 2002, 949 (1-2) :49-60
[9]   A high-performance liquid chromatographic method for screening angiotensin II receptor antagonists in human urine [J].
González, L ;
Alonso, RM ;
Jiménez, RM .
CHROMATOGRAPHIA, 2000, 52 (11-12) :735-740
[10]   Simultaneous determination of hydrochlorothiazide and several angiotensin-II-receptor antagonists by capillary electrophoresis [J].
Hillaert, S ;
Van den Bossche, W .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2003, 31 (02) :329-339