Efficient fractionation and improved protein identification by peptide OFFGEL electrophoresis

被引:173
作者
Hoerth, Patric
Miller, Christine A.
Preckel, Tobias
Wenz, Christian
机构
[1] Agilent Technol Res & Dev & Mkt GMBH & Co KG, D-76337 Waldbronn, Germany
[2] Agilent Technol, Santa Clara, CA 95051 USA
关键词
D O I
10.1074/mcp.T600037-MCP200
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The sample fractionation steps conducted prior to mass detection are critically important for the comprehensive analysis of complex protein mixtures. This paper illustrates the effectiveness of OFFGEL electrophoresis with the Agilent 3100 OFFGEL Fractionator for the fractionation of peptides. An Escherichia coli tryptic digest was separated in 24 fractions, and peptides were identified by reversed-phase liquid chromatography on a microfluidic device with mass spectrometric detection. About 90% of the identified individual peptides were found in only one or two fractions. The distribution of the calculated isoelectric points for the peptides identified in each fraction was especially narrow in the acidic pH range. Standard deviations approached the size of the pH segment covered by the respective fraction. The experimental peptide isoelectric point measured by OFFGEL electrophoresis was used as an additional filter for validation of peptide identifications.
引用
收藏
页码:1968 / 1974
页数:7
相关论文
共 14 条
[1]   ISOELECTRIC-FOCUSING IN IMMOBILIZED PH GRADIENTS - PRINCIPLE, METHODOLOGY AND SOME APPLICATIONS [J].
BJELLQVIST, B ;
EK, K ;
RIGHETTI, PG ;
GIANAZZA, E ;
GORG, A ;
WESTERMEIER, R ;
POSTEL, W .
JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 1982, 6 (04) :317-339
[2]   THE FOCUSING POSITIONS OF POLYPEPTIDES IN IMMOBILIZED PH GRADIENTS CAN BE PREDICTED FROM THEIR AMINO-ACID-SEQUENCES [J].
BJELLQVIST, B ;
HUGHES, GJ ;
PASQUALI, C ;
PAQUET, N ;
RAVIER, F ;
SANCHEZ, JC ;
FRUTIGER, S ;
HOCHSTRASSER, D .
ELECTROPHORESIS, 1993, 14 (10) :1023-1031
[3]   Gel based isoelectric focusing of peptides and the utility of isoelectric point in protein identification [J].
Cargile, BJ ;
Bundy, JL ;
Freeman, TW ;
Stephenson, JL .
JOURNAL OF PROTEOME RESEARCH, 2004, 3 (01) :112-119
[4]   Potential for false positive identifications from large databases through tandem mass spectrometry [J].
Cargile, BJ ;
Bundy, JL ;
Stephenson, JL .
JOURNAL OF PROTEOME RESEARCH, 2004, 3 (05) :1082-1085
[5]   Immobilized pH gradients as a first dimension in shotgun proteomics and analysis of the accuracy of pI predictability of peptides [J].
Cargile, BJ ;
Talley, DL ;
Stephenson, JL .
ELECTROPHORESIS, 2004, 25 (06) :936-945
[6]   A comparison of immobilized pH gradient isoelectric focusing and strong-cation-exchange chromatography as a first dimension in shotgun proteomics [J].
Essader, AS ;
Cargile, BJ ;
Bundy, JL ;
Stephenson, JL .
PROTEOMICS, 2005, 5 (01) :24-34
[7]  
GOERG A, 1985, ELECTROPHORESIS, V6, P599
[8]   Added value for tandem mass spectrometry shotgun proteomics data validation through isoelectric focusing of peptides [J].
Heller, M ;
Ye, ML ;
Michel, PE ;
Morier, P ;
Stalder, D ;
Jünger, MA ;
Aebersold, R ;
Reymond, FR ;
Rossier, JS .
JOURNAL OF PROTEOME RESEARCH, 2005, 4 (06) :2273-2282
[9]   Two-stage Off-Gel™ isoelectric focusing:: Protein followed by peptide fractionation and application to proteome analysis of human plasma [J].
Heller, M ;
Michel, PE ;
Morier, P ;
Crettaz, D ;
Wenz, C ;
Tissot, JD ;
Reymond, F ;
Rossier, JS .
ELECTROPHORESIS, 2005, 26 (06) :1174-1188
[10]   Proteome analysis of human plasma and amniotic fluid by Off-Gel™ isoelectric focusing followed by nano-LC-MS/MS [J].
Michel, PE ;
Crettaz, D ;
Morier, P ;
Heller, M ;
Gallot, D ;
Tissot, JD ;
Reymond, F ;
Rossier, JS .
ELECTROPHORESIS, 2006, 27 (5-6) :1169-1181