Automated hepatic volumetry for living related liver transplantation at multisection CT

被引:140
作者
Nakayama, Yoshiharu
Li, Qiang
Katsuragawa, Shigehiko
Ikeda, Ryuji
Hiai, Yasuhiro
Awai, Kazuo
Kusunoki, Shinichiro
Yamashita, Yasuyuki
Okajima, Hideaki
Inomata, Yukihiro
Doi, Kunio
机构
[1] Kumamoto Univ, Grad Sch Med Sci, Dept Diagnost Radiol, Kumamoto 8608556, Japan
[2] Kumamoto Univ, Grad Sch Med Sci, Dept Transplantat Pediat Surg, Kumamoto 8608556, Japan
[3] Univ Chicago, Kurt Rossmann Labs Radiol Image Res, Dept Radiol, Chicago, IL 60637 USA
[4] Kumamoto Univ, Sch Hlth Sci, Dept Radiol Technol, Kumamoto 8608556, Japan
[5] Kumamoto Univ Hosp, Dept Radiol, Kumamoto, Japan
关键词
D O I
10.1148/radiol.2403050850
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To prospectively compare in vivo hepatic automated volumetry with manual volumetry and measured liver volume. Materials and Methods: The study was conducted in accordance with the guidelines of the Institutional Review Board of Kumamoto University ( Japan). Patient informed consent was obtained. Preoperative multisection computed tomography ( CT) was performed in 35 consecutive patients ( 21 men, 14 women; mean age, 42.8 years; range, 28-72 years) with hepatic disease awaiting living related liver transplantation. The CT scans covered the entire liver at a section thickness of 2.5 mm. Liver volume was estimated by using both the automated and the manual methods. Actual liver weight was obtained for all patients and was converted to hepatic volume on the basis of a predetermined relationship between actual liver weight and volume. Processing time required for both methods was also recorded. Two-tailed paired t test, correlation coefficient, and Bland-Altman tests were used for statistical analyses. Results: Mean liver weight was 881.7 g +/- 249.8 ( standard deviation), and mean measured liver volume was 956.00 cm(3) +/- 280.10. Volumetry performed with the automated and manual methods provided liver volumes of 982.99 cm(3) +/- 301.98 and 937.10 cm(3) +/- 301.31, respectively. There was good correlation between measured and estimated volumes obtained with the automated method ( r = 0.792, P < .01). The manual and automated methods required 32.8 minutes +/- 6.9 and 4.4 minutes +/- 1.9, respectively. Conclusion: The automated method reduced the time required for volumetry of the liver and provided acceptable measurements.
引用
收藏
页码:743 / 748
页数:6
相关论文
共 21 条
[1]   Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Evaluation of living liver donors [J].
Chen, YS ;
Cheng, YF ;
de Villa, VH ;
Wang, CC ;
Lin, CC ;
Huang, TL ;
Jawan, B ;
Chen, CL .
TRANSPLANTATION, 2003, 75 (03) :S16-S19
[4]   DETERMINATION OF LIVER VOLUME FROM CT SCANS USING HISTOGRAM CLUSTER-ANALYSIS [J].
FARJO, LA ;
WILLIAMS, DM ;
BLAND, PH ;
FRANCIS, IR ;
MEYER, CR .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1992, 16 (05) :674-683
[5]   Automatic liver segmentation technique for three-dimensional visualisation of CT data [J].
Gao, LM ;
Heath, DG ;
Kuszyk, BS ;
Fishman, EK .
RADIOLOGY, 1996, 201 (02) :359-364
[6]   MEASUREMENT OF LIVER AND SPLEEN VOLUME BY COMPUTED-TOMOGRAPHY - ASSESSMENT OF REPRODUCIBILITY AND CHANGES FOUND FOLLOWING A SELECTIVE DISTAL SPLENORENAL SHUNT [J].
HENDERSON, JM ;
HEYMSFIELD, SB ;
HOROWITZ, J ;
KUTNER, MH .
RADIOLOGY, 1981, 141 (02) :525-527
[7]   Liver segmentation in living liver transplant donors: Comparison of semiautomatic and manual methods [J].
Hermoye, L ;
Laamari-Azjal, I ;
Cao, ZJ ;
Annet, L ;
Lerut, J ;
Dawant, BM ;
Van Beers, BE .
RADIOLOGY, 2005, 234 (01) :171-178
[8]   ACCURATE MEASUREMENT OF LIVER, KIDNEY, AND SPLEEN VOLUME AND MASS BY COMPUTERIZED AXIAL-TOMOGRAPHY [J].
HEYMSFIELD, SB ;
FULENWIDER, T ;
NORDLINGER, B ;
BARLOW, R ;
SONES, P ;
KUTNER, M .
ANNALS OF INTERNAL MEDICINE, 1979, 90 (02) :185-187
[9]   Fulminant hepatic failure: Observation with serial CT [J].
Itai, Y ;
Sekiyama, K ;
Ahmadi, T ;
Obuchi, M ;
Yoshiba, M .
RADIOLOGY, 1997, 202 (02) :379-382
[10]   Accuracy of volumetric measurements after virtual right hepatectomy in potential donors undergoing living adult liver transplantation [J].
Kamel, IR ;
Kruskal, JB ;
Warmbrand, G ;
Goldberg, SN ;
Pomfret, EA ;
Raptopoulos, V .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :483-487