Sensitivity in Type 1 Diabetic Subjects Wearing a SensorAugmented Insulin Pump

被引:41
作者
Schiavon, Michele [1 ]
Man, Chiara Dalla [1 ]
Kudva, Yogish C. [2 ]
Basu, Ananda [2 ]
Cobelli, Claudio [1 ]
机构
[1] Univ Padua, Dept Informat Engn, Padua, Italy
[2] Mayo Clin & Mayo Fdn, Dept Internal Med, Div Endocrinol Diabet Metab & Nutr, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
MINIMAL MODEL; GLUCOSE-ABSORPTION; ORAL-TEST; VALIDATION; SECRETION; CLAMP;
D O I
10.2337/dc13-1120
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVEThe goal was to develop a new index of insulin sensitivity in patients with type 1 diabetes estimated from continuous glucose monitoring (CGM) and subcutaneous insulin delivery data under carefully controlled conditions.RESEARCH DESIGN AND METHODSThe database consists of 12 subjects with type 1 diabetes, studied during breakfast, lunch, and dinner, in a clinical research unit, wearing both subcutaneous insulin pump and CGM device. Frequent blood samples were drawn for measurements of plasma glucose and insulin concentrations in order to estimate insulin sensitivity with the oral minimal model (S-I(MM)). The new index of insulin sensitivity (S-I(SP)) was calculated with a simple algebraic formula for each meal, using only CGM and insulin pump data and compared with S-I(MM).RESULTSSISP was well correlated with S-I(MM) (r = 0.825; P < 10(-8)), and diurnal pattern was also similar to S-I(MM).CONCLUSIONSA novel method for estimating insulin sensitivity in subjects with type 1 diabetes on sensor-augmented insulin pump therapy has been presented. This new index correlates well with the reference oral minimal model estimate of insulin sensitivity. The knowledge of patient-specific insulin sensitivity and its diurnal variation can help in optimizing insulin therapy in type 1 diabetes and could also inform next-generation closed-loop control systems.
引用
收藏
页码:1216 / 1223
页数:8
相关论文
共 24 条
[1]   Effects of Type 2 Diabetes on Insulin Secretion, Insulin Action, Glucose Effectiveness, and Postprandial Glucose Metabolism [J].
Basu, Ananda ;
Man, Chiara Dalla ;
Basu, Rita ;
Toffolo, Gianna ;
Cobelli, Claudio ;
Rizza, Robert A. .
DIABETES CARE, 2009, 32 (05) :866-872
[2]   QUANTITATIVE ESTIMATION OF INSULIN SENSITIVITY [J].
BERGMAN, RN ;
IDER, YZ ;
BOWDEN, CR ;
COBELLI, C .
AMERICAN JOURNAL OF PHYSIOLOGY, 1979, 236 (06) :E667-E677
[3]   Calculatin the insulin to carbohydrate ratio using the hyperinsulinaemic-euglycaemic clamp - a novel use for a proven technique [J].
Bevier, Wendy C. ;
Zisser, Howard ;
Palerm, Cesar C. ;
Finan, Daniel A. ;
Seborg, Dale E. ;
Doyle, Francis J., III ;
Wollitzer, Alison Okada ;
Jovanovic, Lois .
DIABETES-METABOLISM RESEARCH AND REVIEWS, 2007, 23 (06) :472-478
[4]  
Breton MD, 2008, PCT international patent application PCT/US2008/069416, Patent No. 2008069416
[5]   Minimal model assessment of hepatic insulin extraction during an oral test from standard insulin kinetic parameters [J].
Campioni, M. ;
Toffolo, G. ;
Basu, R. ;
Rizza, R. A. ;
Cobelli, C. .
AMERICAN JOURNAL OF PHYSIOLOGY-ENDOCRINOLOGY AND METABOLISM, 2009, 297 (04) :E941-E948
[6]   Insulin sensitivity from meal tolerance tests in normal subjects: A minimal model index [J].
Caumo, A ;
Bergman, RN ;
Cobelli, C .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2000, 85 (11) :4396-4402
[7]  
Cobelli C, 2012, U.S. patent application US, Patent No. [13/661,755, 13661755]
[8]   Insulin sensitivity by oral glucose minimal models: validation against clamp [J].
Dalla Man, C ;
Yarasheski, KE ;
Caumo, A ;
Robertson, H ;
Toffolo, G ;
Polonsky, KS ;
Cobelli, C .
AMERICAN JOURNAL OF PHYSIOLOGY-ENDOCRINOLOGY AND METABOLISM, 2005, 289 (06) :E954-E959
[9]   Minimal model estimation of glucose absorption and insulin sensitivity from oral test: validation with a tracer method [J].
Dalla Man, C ;
Caumo, A ;
Basu, R ;
Rizza, R ;
Toffolo, G ;
Cobelli, C .
AMERICAN JOURNAL OF PHYSIOLOGY-ENDOCRINOLOGY AND METABOLISM, 2004, 287 (04) :E637-E643
[10]   A system model of oral glucose absorption: Validation on gold standard data [J].
Dalla Man, Chiara ;
Camilleri, Michael ;
Cobelli, Claudio .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (12) :2472-2478