Failure patterns of four osseointegrated oral implant systems

被引:47
作者
Esposito, M
Hirsch, JM
Lekholm, U
Thomsen, P
机构
[1] GOTHENBURG UNIV,DEPT ORAL & MAXILLOFACIAL SURG,S-41390 GOTHENBURG,SWEDEN
[2] GOTHENBURG UNIV,FAC ODONTOL,BRANEMARK CLIN,S-41390 GOTHENBURG,SWEDEN
[3] UNIV UPPSALA HOSP,DEPT ORAL & MAXILLOFACIAL SURG,S-75185 UPPSALA,SWEDEN
关键词
D O I
10.1023/A:1018593403126
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of th is metanalysis was to investigate possible differences in failure patterns among four different osseointegrated ora I implant systems. On ly systems with a supposed scientific validation, based on long-term follow-up studies, were selected for this analysis, i.e. the Branemark system, the Calcitek Integral system, the Interpore IMZ system and the Straumann ITI system. While several prospective reports could be found on the Branemark system, only a few retrospective investigations concerning the ITI and the IMZ systems were available. No data on the Integral system could be employed. Despite these limitations, a substantial difference in failure patterns among various implant systems was observed. The Branemark implants showed a higher incidence of early failures, though sharply decreasing over time. On the contrary, IMZ implant characterized by rougher surfaces displayed a lower incidence of early failures, but showed constant or increased failure rates over time. For the ITI im pla nts, for example, a higher prevalence of late failures, attributable to chronic bacterial infection (peri-implantitis), was observed when compared to the Branemark system. With the exception of the latter implant system, prospective long-term follow-up studies, using similar and well-defined success criteria, are needed for the others to confirm the current preliminary findings.
引用
收藏
页码:843 / 847
页数:5
相关论文
共 42 条
[1]   A 15-YEAR STUDY OF OSSEOINTEGRATED IMPLANTS IN THE TREATMENT OF THE EDENTULOUS JAW [J].
ADELL, R ;
LEKHOLM, U ;
ROCKLER, B ;
BRANEMARK, PI .
INTERNATIONAL JOURNAL OF ORAL SURGERY, 1981, 10 (06) :387-416
[2]  
ASTRAND P, 1991, International Journal of Oral and Maxillofacial Implants, V6, P305
[3]   Non-submerged implants in the treatment of the edentulous lower jaw - A 2-year longitudinal study [J].
Astrand, P ;
Almfeldt, I ;
Brunell, G ;
Hamp, SE ;
Hellem, S ;
Karlsson, U .
CLINICAL ORAL IMPLANTS RESEARCH, 1996, 7 (04) :337-344
[4]   TITANIUM PLASMA-SPRAYED (TPS) SCREW IMPLANTS FOR THE RECONSTRUCTION OF THE EDENTULOUS MANDIBLE [J].
BABBUSH, CA ;
KENT, JN ;
MISIEK, DJ .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1986, 44 (04) :274-282
[5]   LONG-TERM FOLLOW-UP ON HYDROXYLAPATITE-COATED CYLINDRICAL DENTAL IMPLANTS - A COMPARISON BETWEEN DEVELOPMENTAL AND RECENT PERIODS [J].
BLOCK, MS ;
KENT, JN .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1994, 52 (09) :937-943
[6]  
BRANEMARK P I, 1969, Scandinavian Journal of Plastic and Reconstructive Surgery, V3, P81
[7]   Implant-retained mandibular overdentures with immediate loading - A retrospective multicenter study on 226 consecutive cases [J].
Chiapasco, M ;
Gatti, C ;
Rossi, E ;
Haefliger, W ;
Markwalder, TH .
CLINICAL ORAL IMPLANTS RESEARCH, 1997, 8 (01) :48-57
[8]  
d'Hoedt B, 1989, Int J Oral Maxillofac Implants, V4, P95
[9]  
DEPORTER DA, 1987, INT J ORAL IMPLANTOL, V4, P31
[10]   FOOD-AND-DRUG-ADMINISTRATION REQUIREMENTS FOR DENTAL IMPLANTS [J].
ECKERT, SE .
JOURNAL OF PROSTHETIC DENTISTRY, 1995, 74 (02) :162-168