Commissioning and initial operation of a radioactive beam ion source at ISAC

被引:66
作者
Dombsky, M [1 ]
Bishop, D [1 ]
Bricault, P [1 ]
Dale, D [1 ]
Hurst, A [1 ]
Jayamanna, K [1 ]
Keitel, R [1 ]
Olivo, M [1 ]
Schmor, P [1 ]
Stanford, G [1 ]
机构
[1] TRIUMF, Vancouver, BC V6T 2A3, Canada
关键词
D O I
10.1063/1.1150364
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In November of 1998, the ISAC radioactive beam facility at TRIUMF started delivering on-line isotope separated radioactive beams to experiments. A surface ionization source developed for ISAC has been used to commission the mass separator and beam transport systems and is providing radioactive beams to the first generation of ISAC experiments. The ion source is integral with the radioactive beam production target and is designed to be simple, radiation hard, inexpensive, and easily exchanged by remote-handling techniques. The ion source and its extraction column are suspended at the bottom of similar to 2 m of steel shielding incorporated in the target module. The module is suspended in a vacuum tank with primary and secondary vacuum systems. All services for the target/ion source and beam extraction system are ducted through the module shielding. The first sets of beam transport elements and beam diagnostic devices are similarly suspended in vacuum at the bottom of two additional shielded modules. Ion beam characteristics can be routinely monitored during on-line operation by a system of Faraday cups, wire scanners, "harp" monitors, and a novel emittance measurement apparatus that can measure beam emittance in both horizontal and vertical planes. The diagnostics devices are capable of resolving beam signals down to the 10 pA range. (C) 2000 American Institute of Physics. [S0034-6748(00)64102-3].
引用
收藏
页码:978 / 980
页数:3
相关论文
共 4 条
[1]   AN EMITTANCE SCANNER FOR INTENSE LOW-ENERGY ION-BEAMS [J].
ALLISON, PW ;
SHERMAN, JD ;
HOLTKAMP, DB .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1983, 30 (04) :2204-2206
[2]  
Bricault P. G., 1997, P 15 INT C CYCL THEI, VB126, P213
[3]   Evaluation of a prototype Isotope Separator Accelerator surface ionization source [J].
Dombsky, M ;
Baartman, R ;
Bricault, P ;
Doornbos, J ;
Jayamanna, K ;
Kuo, T ;
Mackenzie, G ;
McDonald, M ;
Schmor, P ;
Yuan, D .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (02) :1170-1172
[4]   An ion source test stand for the ISAC facility at TRIUMF [J].
Dombsky, M ;
Baartman, R ;
Doornbos, J ;
Hodges, T ;
Jayamanna, K ;
Keitel, R ;
Kuo, T ;
Mackenzie, G ;
McDonald, M ;
Schmor, P ;
Yin, Y .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1997, 126 (1-4) :50-54