Determination of sixteen UV filters in suncare formulations by high-performance liquid chromatography

被引:49
作者
Schakel, DJ [1 ]
Kalsbeek, D [1 ]
Boer, K [1 ]
机构
[1] Food & Consumer Prod Safety Author, Insepectrorate Hlth Protect & Vet Publ Hlth, VWA KvW, Res Lab Non Food Chem, NL-9726 BA Groningen, Netherlands
关键词
suncare products; UV filters; tween; 80; EDTA;
D O I
10.1016/j.chroma.2004.07.068
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Reversed-phase liquid chromatography (RP-LC) with UV detection was developed for the simultaneous determination of 16 organic UV filters worldwide authorised in suncare products. The filters determined were: 4-aminobenzoic acid, homosalate, benzophenone-3,2-phenylbenzimidazole-5-sulfonic acid, terephthalidene dicamphor sulfonic acid, 4-tert-butyl-4'-methoxy-dibenzoylmethane, octocrylene, 2-ethythexyl-4-methoxycinnamate, isoamyl-p-methoxycinnamate, ethylhexyltriazone, drometrizole trisiloxane, diethylhexyl butamido triazone, 3-(4-methylbenzyliden) camphor, 2-ethylhexylsalicylate, 2-ethylhexyl-4-dimethylaminobenzoate and benzophenone-4. A C-18 stationary phase and a gradient of ethanol-aqueous acetate buffer containing 0.2 mM of EDTA, was used with a flow-rate of 1.0 ml/mm. UV detection was carried out at 313 and 360 nm. The analysis required 32 min and the limits of detection were between 30 and 4130 mg/kg in the original suncare product. Tween 80 was used to break down the different emulsions in order to procure a proper extraction of the UV filters. The method was validated for UV filters in three matrices, oil, water-in-oil emulsion and oil-in-water emulsion. Recoveries from spiked samples were 86-113% depending on the matrix used. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:127 / 130
页数:4
相关论文
共 9 条
[1]   Determination of the UV filters worldwide authorised in sunscreens by high-performance liquid chromatography -: Use of cyclodextrins as mobile phase modifier [J].
Chisvert, A ;
Pascual-Martí, MC ;
Salvador, A .
JOURNAL OF CHROMATOGRAPHY A, 2001, 921 (02) :207-215
[2]  
FRANKE I, 2003, QUALITY ASSURANCE HD
[3]  
Hauri Urs, 2003, Mitteilungen aus Lebensmitteluntersuchung und Hygiene, V94, P80
[4]  
MAJORS RE, LC GC JUL
[6]   Identification of UV filters in sunscreen products by high-performance liquid chromatography diode-array detection [J].
Rastogi, SC ;
Jensen, GH .
JOURNAL OF CHROMATOGRAPHY A, 1998, 828 (1-2) :311-316
[7]  
Schneider P., 1996, DRUG COSMET IND, V159, P32
[8]  
STERN A, 2002, QUALITY ASSURANCE HD
[9]  
1995, AOAC PEER VERIFIED M