Effect of skin impedance on image quality and variability in electrical impedance tomography: A model study

被引:41
作者
Boone, KG
Holder, DS
机构
[1] Department of Physiology, University College London, London WC1E 6BT, Gower Street
[2] University College Hospital, London
[3] University of California, Berkeley, CA
[4] St. James's University Hospital, Leeds
[5] Department of Physiology, University College, London
[6] Royal Brompton Hospital, London
关键词
electrical impedance tomography; impedance; skin;
D O I
10.1007/BF02520003
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A computer simulation is used to investigate the relationship between skin impedance and image artefacts in electrical impedance tomography. Sets of electrode impedance are generated with a pseudo-random distribution and used to introduce errors in boundary voltage measurements. To simplify the analysis, the non-idealities in the current injection circuit are replaced by a fixed common-mode error term. The boundary voltages are reconstructed into images and inspected. Where the simulated skin impedance remains constant between measurements, large impedances (>2k Omega) do not cause significant degradation of the image. Where the skin impedances 'drift' between measurements, a drift of 5% from a starting impedance of 100 Omega is sufficient to cause significant image distortion. If the skin impedances vary randomly between measurements; they have to be less than 10 Omega to allow satisfactory images. Skin impedances' are typically 100-200 Omega at 50 kHz on unprepared skin. These values are sufficient to cause Image distortion if they drift over time. It is concluded that the patient's skin should be abraded to reduce impedance, and measurements should be avoided in the first 10 min after electrode placement.
引用
收藏
页码:351 / 354
页数:4
相关论文
共 10 条
  • [1] BARBER DC, 1987, CLIN PHYS PHYSL M SA, V8, P91
  • [2] BOONE K, 1995, INNOV TECH BIOL MED, V16, P49
  • [3] BOONE KG, 1995, INNOV TECH BIOL MED, V16, P611
  • [4] Geddes LA, 1972, ELECTRODES MEASUREME
  • [5] FINITE-ELEMENT MODELING OF ELECTRODE SKIN CONTACT IMPEDANCE IN ELECTRICAL-IMPEDANCE TOMOGRAPHY
    HUA, P
    WOO, EJ
    WEBSTER, JG
    TOMPKINS, WJ
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (04) : 335 - 343
  • [6] USING COMPOUND ELECTRODES IN ELECTRICAL-IMPEDANCE TOMOGRAPHY
    HUA, P
    WOO, EJ
    WEBSTER, JG
    TOMPKINS, WJ
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (01) : 29 - 34
  • [7] MCADAMS ET, 1990, P 3 EUR COMM WORKSH, P14
  • [8] RECORD PM, 1990, P COP M EL IMP TOM, P168
  • [9] RIU PJ, 1990, P COP M EL IMP TOM S, P198
  • [10] WOO EJ, 1990, THESIS U WISCONSIN M