Accuracy of glucose reflectance testing for detecting hypoglycemia in term newborns

被引:9
作者
Meloy, L [1 ]
Miller, G [1 ]
Chandrasekaran, MH [1 ]
Summitt, G [1 ]
Gutcher, G [1 ]
机构
[1] Virginia Commonwealth Univ, Med Coll Virginia, Richmond, VA 23298 USA
关键词
D O I
10.1177/000992289903801204
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Glucose concentrations of 248 newborns at 1 and 3 hours of age were determined at the bedside by means of the Accu-Chek III Reaectance(R) meter and in the laboratory by means of the Vitros 750(R) analyzer, Hypoglycemia being defined as a glucose concentration less than 40 mg/dL, the reflectance meter had a sensitivity of 0.757 and a specificity of 0.876 at 1 hour and 0.975 at 3 hours. The imprecision of the reflectance meter at 1 hour eliminated it as a routine testing method.
引用
收藏
页码:717 / 724
页数:8
相关论文
共 12 条
[1]  
COLE P, 1994, J PERINAT, V14, P118
[2]   CLINICAL-APPLICATION OF A NEW GLUCOSE ANALYZER IN THE NEONATAL INTENSIVE-CARE UNIT - COMPARISON WITH OTHER METHODS [J].
CONRAD, PD ;
SPARKS, JW ;
OSBERG, I ;
ABRAMS, L ;
HAY, WW .
JOURNAL OF PEDIATRICS, 1989, 114 (02) :281-287
[3]  
CORNBLATH M, 1993, J PEDIATR ENDOCRINOL, V6, P113
[4]  
CORNBLATH M, 1976, DISORDERS CARBOHYDRA, P187
[5]  
CORNBLEET PJ, 1979, CLIN CHEM, V25, P432
[6]   Neonatal hypoglycemia, part II: Pathophysiology and therapy [J].
Halamek, LP ;
Stevenson, DK .
CLINICAL PEDIATRICS, 1998, 37 (01) :11-16
[7]  
MERENSTEIN GB, 1993, PEDIATRICS, V92, P474
[8]   BASIC PRINCIPLES OF ROC ANALYSIS [J].
METZ, CE .
SEMINARS IN NUCLEAR MEDICINE, 1978, 8 (04) :283-298
[9]  
NICOLS J, 1995, AM J CLIN PATHOL, V103, P244
[10]   Laboratory and clinical evaluation of two glucose meters for the neonatal intensive care unit [J].
Perkins, SL ;
Doelle, H ;
Escares, E ;
Forsythe, J ;
Pronovost, C ;
Taylor-Clapp, S .
CLINICAL BIOCHEMISTRY, 1998, 31 (02) :67-71