摘要
The results of a quantitative study of the impact of sound speed errors on the spatial resolution and amplitude sensitivity of a commercial medical ultrasound scanner are presented in the context of their clinical significance. The beamforming parameters of the scanner were manipulated to produce sound speed errors ranging over +/-8% while imaging a wire target and an attenuating, speckle-generating phantom. For the wire target, these errors produced increases in lateral beam width of up to 320% and reductions in peak echo amplitude of up to 10.5 dB. In the speckle-generating phantom, these errors produced increases in speckle intensity correlation cell area of up to 92% and reductions in mean speckle brightness of up to 5.6 dB. These results are applied in statistical analyses of two detection tasks of clinical relevance. The first is of low contrast lesion detectability, predicting the changes in the correct decision probability as a function of lesion size, contrast, and sound speed error. The second is of point target detectability, predicting the changes in the correct decision probability as function of point target reflectivity and sound speed error. Representative results of these analyses are presented and their implications for clinical imaging are discussed. In general, sound speed errors have a more significant impact on point target detectability over lesion detectability by these analyses, producing up to a 22% reduction in correct decisions for a typical error. (C) 2000 Acoustical Society of America. [S0001-4966(00)01206-6].
引用
收藏
页码:3540 / 3548
页数:9
相关论文
共 35 条
[1]   The direct estimation of sound speed using pulse-echo ultrasound [J].
Anderson, ME ;
Trahey, GE .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1998, 104 (05) :3099-3106
[2]   The detection of breast microcalcifications with medical ultrasound [J].
Anderson, ME ;
Soo, MS ;
Bentley, RC ;
Trahey, GE .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1997, 101 (01) :29-39
[3]   SPEED OF SOUND IN PURE WATER AS A FUNCTION OF TEMPERATURE [J].
BILANIUK, N ;
WONG, GSK .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1993, 93 (03) :1609-1612
[4]  
Bracewell Ronald N, 1986, FOURIER TRANSFORM IT, V31999
[5]   PHASE-ABERRATION CORRECTION USING SIGNALS FROM POINT REFLECTORS AND DIFFUSE SCATTERERS - BASIC PRINCIPLES [J].
FLAX, SW ;
ODONNELL, M .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (06) :758-767
[6]  
GOODMAN JW, 1985, STAT OPTICS, P44
[7]  
GOODMAN JW, 1984, LASER SPECKLE RELATE, P36
[8]   COMPILATION OF EMPIRICAL ULTRASONIC PROPERTIES OF MAMMALIAN-TISSUES .2. [J].
GOSS, SA ;
JOHNSTON, RL ;
DUNN, F .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1980, 68 (01) :93-108
[9]   COMPREHENSIVE COMPILATION OF EMPIRICAL ULTRASONIC PROPERTIES OF MAMMALIAN-TISSUES [J].
GOSS, SA ;
JOHNSTON, RL ;
DUNN, F .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1978, 64 (02) :423-457
[10]   ULTRASOUND CONTRAST-DETAIL ANALYSIS - A PRELIMINARY-STUDY IN HUMAN OBSERVER PERFORMANCE [J].
HALL, TJ ;
INSANA, MF ;
SOLLER, NM ;
HARRISON, LA .
MEDICAL PHYSICS, 1993, 20 (01) :117-127