Randomly segmented central k-space ordering in high-spatial-resolution contrast-enhanced MR angiography of the supraaortic arteries: Initial experience

被引:107
作者
Willinek, WA
Gieseke, J
Conrad, R
Strunk, H
Hoogeveen, R
von Falkenhausen, M
Keller, E
Urbach, H
Kuhl, CK
Schild, HH
机构
[1] Univ Bonn, Dept Radiol, D-53105 Bonn, Germany
[2] Philips Med Syst, Best, Netherlands
关键词
arteries; basilar; stenosis or obstruction; carotid arteries; magnetic resonance (MR); k-space; vascular studies;
D O I
10.1148/radiol.2252011167
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Contrast material-enhanced three-dimensional (3D) magnetic resonance (MR) angiography of the supraaortic arteries with randomly segmented central k-space ordering (ie, contrast-enhanced timing-robust angiography [CENTRA]) was performed in 16 patients. CENTRA enabled reliable depiction of the aortic arch up to the circle of Willis at high spatial resolution (true voxel size, 0.81 X 0.81 X 1.0 mm(3)). With CENTRA, the divergent demands of, high spatial resolution, wide anatomic coverage, arterial phase imaging have been reconciled. The random order of central k-space acquisition may minimize artifacts in contrast-enhanced 3D MR,angiography caused by unstable contrast material opacification at the initiation of sampling. (C) RSNA, 2002.
引用
收藏
页码:583 / 588
页数:6
相关论文
共 13 条
[1]   Contrast-enhanced carotid MR angiography with commercially available triggering mechanisms and elliptic centric phase encoding [J].
De Marco, JK ;
Schonfeld, S ;
Keller, I ;
Bernstein, MA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (01) :221-227
[2]  
Fain SB, 1999, MAGNET RESON MED, V42, P1106, DOI 10.1002/(SICI)1522-2594(199912)42:6<1106::AID-MRM15>3.0.CO
[3]  
2-Q
[4]   Fluoroscopically triggered contrast-enhanced 3D MR DSA and 3D time-of-flight turbo MRA of the carotid arteries: first clinical experiences in correlation with ultrasound, x-ray angiography, and endarterectomy findings [J].
Fellner, FA ;
Fellner, C ;
Wutke, R ;
Lang, W ;
Laub, G ;
Schmidt, M ;
Janka, R ;
Denzel, C ;
Bautz, W .
MAGNETIC RESONANCE IMAGING, 2000, 18 (05) :575-585
[5]   HOW TO MEASURE CAROTID STENOSIS [J].
FOX, AJ .
RADIOLOGY, 1993, 186 (02) :316-318
[6]  
Harvey P, 2000, International patent, Patent No. [PCT/EP01/02907, 0102907]
[7]   Carotid arteries: Maximizing arterial to venous contrast in fluoroscopically triggered contrast-enhanced MR angiography with elliptic centric view ordering [J].
Huston, J ;
Fain, SB ;
Riederer, SJ ;
Wilman, AH ;
Bernstein, MA ;
Busse, RF .
RADIOLOGY, 1999, 211 (01) :265-273
[8]  
Huston J, 2001, RADIOLOGY, V218, P138
[9]  
Leclerc X, 2000, AM J NEURORADIOL, V21, P1021
[10]   BREATH-HOLD GADOLINIUM-ENHANCED MR-ANGIOGRAPHY OF THE ABDOMINAL-AORTA AND ITS MAJOR BRANCHES [J].
PRINCE, MR ;
NARASIMHAM, DL ;
STANLEY, JC ;
CHENEVERT, TL ;
WILLIAMS, DM ;
MARX, V ;
CHO, KJ .
RADIOLOGY, 1995, 197 (03) :785-792