Comparison of calibration methods for atomic-force microscopy cantilevers

被引:312
作者
Burnham, NA [1 ]
Chen, X
Hodges, CS
Matei, GA
Thoreson, EJ
Roberts, CJ
Davies, MC
Tendler, SJB
机构
[1] Worcester Polytech Inst, Dept Phys, Worcester, MA 01609 USA
[2] Univ Nottingham, Sch Pharmaceut Sci, Lab Biophys & Surface Anal, Nottingham NG7 2RD, England
[3] Univ Leeds, Dept Chem Engn, Leeds LS2 9JT, W Yorkshire, England
关键词
D O I
10.1088/0957-4484/14/1/301
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
The scientific community needs a rapid and reliable way of accurately determining the stiffness of atomic-force microscopy cantilevers. We have compared the experimentally determined values of stiffness for ten cantilever probes using four different methods. For rectangular silicon cantilever beams of well defined geometry, the approaches all yield values within 17% of the manufacturer's nominal stiffness. One of the methods is new, based on the acquisition and analysis of thermal distribution functions of the oscillator's amplitude fluctuations. We evaluate this method in comparison to the three others and recommend it for its ease of use and broad applicability.
引用
收藏
页码:1 / 6
页数:6
相关论文
共 23 条
  • [1] [Anonymous], 1962, NOISE FLUCTUATIONS I
  • [2] ATKINS PW, 1978, PHYSICAL CHEM
  • [3] CALCULATION OF THERMAL NOISE IN ATOMIC-FORCE MICROSCOPY
    BUTT, HJ
    JASCHKE, M
    [J]. NANOTECHNOLOGY, 1995, 6 (01) : 1 - 7
  • [4] SCAN SPEED LIMIT IN ATOMIC FORCE MICROSCOPY
    BUTT, HJ
    SIEDLE, P
    SEIFERT, K
    FENDLER, K
    SEEGER, T
    BAMBERG, E
    WEISENHORN, AL
    GOLDIE, K
    ENGEL, A
    [J]. JOURNAL OF MICROSCOPY-OXFORD, 1993, 169 : 75 - 84
  • [5] Experimental validation of theoretical models for the frequency response of atomic force microscope cantilever beams immersed in fluids
    Chon, JWM
    Mulvaney, P
    Sader, JE
    [J]. JOURNAL OF APPLIED PHYSICS, 2000, 87 (08) : 3978 - 3988
  • [6] A NONDESTRUCTIVE METHOD FOR DETERMINING THE SPRING CONSTANT OF CANTILEVERS FOR SCANNING FORCE MICROSCOPY
    CLEVELAND, JP
    MANNE, S
    BOCEK, D
    HANSMA, PK
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (02) : 403 - 405
  • [7] Determination of the spring constants of probes for force microscopy/spectroscopy
    Gibson, CT
    Watson, GS
    Myhra, S
    [J]. NANOTECHNOLOGY, 1996, 7 (03) : 259 - 262
  • [8] HAZEL J, 1996, POLYM PREPR, V37, P567
  • [9] Friction force microscopy measurements: Normal and torsional spring constants for V-shaped cantilevers
    Hazel, JL
    Tsukruk, VV
    [J]. JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 1998, 120 (04): : 814 - 819
  • [10] Spring constants of composite ceramic/gold cantilevers for scanning probe microscopy
    Hazel, JL
    Tsukruk, VV
    [J]. THIN SOLID FILMS, 1999, 339 (1-2) : 249 - 257