Determination of tin in steels by non-dispersive atomic fluorescence spectrometry coupled with flow-injection hydride generation in the presence of L-cysteine

被引:21
作者
Chen, HW
Yao, W
Wu, DX
Brindle, ID
机构
[1] BROCK UNIV,DEPT CHEM,ST CATHARINES,ON L2S 3A1,CANADA
[2] HANGZHOU UNIV,DEPT CHEM,HANGZHOU 310028,ZHEJIANG,PEOPLES R CHINA
[3] HANGZHOU UNIV,DEPT GEOG,HANGZHOU 310028,ZHEJIANG,PEOPLES R CHINA
关键词
atomic fluorescence; L-cysteine; flow injection; hydride generation; tin;
D O I
10.1016/S0584-8547(96)01563-7
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Flow-injection hydride generation has been coupled with a non-dispersive atomic fluorescence spectrometer for the determination of trace concentrations of tin. The use of L-cysteine in stannane generation enhances the atomic fluorescence signal for tin. In addition, it appears that the how-injection method allows the use of a wider range of acid concentrations, and hence interference from transition elements is also reduced. An absolute detection limit of 1.3 ng was achieved with a 0.5 mi sample. Ten determinations of a solution containing 100 ng ml(-1) of tin(H) generated a relative standard deviation of 1.8%. The linear dynamic range of the calibration curve extended over three orders of magnitude and sample throughput rate was 144 samples h(-1). Application of the proposed technique is demonstrated by the determination of tin in low alloy steel standard reference materials.
引用
收藏
页码:1829 / 1836
页数:8
相关论文
共 25 条