Background and hypothesis: This study was carried out to determine whether cardiac troponin T test in rapid assay gives positive results in patients previously submitted to cardioversion or electrical defibrillation. Methods: Forty patients with supraventricular tachyarrhythmias lasting no more than 2 days were treated with electrical cardioversion. The total creatine phosphokinase (CPK)-MB isoenzyme and troponin T in rapid assay were measured at baseline and at 6, 12, and 24 h thereafter. Results: Total CPK baseline levels were normal in all cases; within 4 h, the serum CPK levels increased by 98%, at 6 h by 111.5%, at 12 h by 168%, and at 24 h by 225% (p > 0.01). The CPK-MB isoenzyme showed no percentage increase of total CPK higher than 5%, measured at 6, 12, and 24 h after the shock, independent of the number of attempts of cardioversion. The troponin T test was also negative in all cases at baseline and at 6, 12, and 24 h after cardioversion. Conclusion: We conclude that the absence of elevations in CPK-MB levels and cardiac troponin T levels matched clinical and electrocardiographic results showing absence of myocardial damage after electrical cardioversion.
引用
收藏
页码:619 / 621
页数:3
相关论文
共 27 条
[1]   USE OF EXTERNAL ELECTRIC COUNTERSHOCK IN TREATMENT OF VENTRICULAR TACHYCARDIA [J].
ALEXANDER, S ;
KLEIGER, R ;
LOWN, B .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1961, 177 (13) :916-&
[2]  
Clarence Davies III J., 1975, P CARDIAC DEFIBRILLA, P27
[3]   CARDIAC INOTROPIC AND CORONARY VASCULAR RESPONSES TO COUNTERSHOCK - EVIDENCE FOR EXCITATION OF INTRACARDIAC NERVES [J].
COBB, FR ;
WALLACE, AG ;
WAGNER, GS .
CIRCULATION RESEARCH, 1968, 23 (06) :731-&
[4]  
COLLINSON PO, 1992, NEW ENGL J MED, V327, P1760
[5]  
CORBITT D, 1969, AM J CLIN PATHOL, V512, P107
[6]   MYOCARDIAL NECROSIS FROM DIRECT-CURRENT COUNTERSHOCK - EFFECT OF PADDLE ELECTRODE SIZE AND TIME INTERVAL BETWEEN DISCHARGES [J].
DAHL, CF ;
EWY, GA ;
WARNER, ED ;
THOMAS, ED .
CIRCULATION, 1974, 50 (05) :956-961
[7]   CARDIOVERSION AND DEFIBRILLATION [J].
DESILVA, RA ;
GRABOYS, TB ;
PODRID, PJ ;
LOWN, B .
AMERICAN HEART JOURNAL, 1980, 100 (06) :881-895
[8]  
EHSANI A, 1976, AM J CARDIOL, V12, P37
[9]   CONTROLLED NUCLEATION FOR REGULATION OF PARTICLE-SIZE IN MONODISPERSE GOLD SUSPENSIONS [J].
FRENS, G .
NATURE-PHYSICAL SCIENCE, 1973, 241 (105) :20-22
[10]   TRANSIENT MYOCARDIAL INJURY AFTER ELECTIVE ELECTRICAL CARDIOVERSION [J].
GARCIARUBIRA, JC ;
ROMERO, D ;
GARCIA, JT ;
LOPEZ, V ;
CRUZ, JM .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1994, 46 (03) :283-285