A fast and reliable route integrating calibration and analysis protocols for water-soluble vitamin determination on microchip-electrochemistry platforms

被引:17
作者
Gonzalez Crevillen, Agustin [1 ]
Javier Blasco, Antonio [1 ]
Cristina Gonzalez, Maria [1 ]
Escarpa, Alberto [1 ]
机构
[1] Univ Alcala de Henares, Dpto Quim Analit & Ingn Quim, E-28871 Alcala De Henares, Spain
关键词
calibration; CE microchip; electrochemical detection; water-soluble vitamins;
D O I
10.1002/elps.200600213
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel analytical route to determine water-soluble vitamins (B group and C) using single channel microchip-electrochemistry platforms is presented. The electrochemical detection protocol was carefully optimized, and it was shown that it was crucial to use 1 M nitric acid in the detector compartment to detect folic acid. A phosphate buffer (pH 6, 10 mM) and a separation voltage of 2 kV gave the complete separation of vitamins in less than 130 s, with good reproducibility (RSDs less than 10%) and accuracy (error less than 9%). In addition, a methodological innovation integrating calibration and analysis of water-soluble vitamins on the chip is also proposed. The strategy consisted in sequentially using both reservoirs (named calibration and analysis reservoirs) as well as a calibration factor (defined as signal/concentration of analyte). The analytical route required 350 s in the overall protocol (employing 130 s in calibration plus 130 s in analysis), an improvement over the times used in both conventional and microchip protocols.
引用
收藏
页码:5110 / 5118
页数:9
相关论文
共 58 条
[1]  
BLASCO A, 2005, ELECTROCHEMICAL DETE, P703
[2]   Fast and simultaneous detection of prominent natural antioxidants using analytical microsystems for capillary electrophoresis with a glassy carbon electrode:: A new gateway to food environments [J].
Blasco, AJ ;
Barrigas, I ;
González, MC ;
Escarpa, A .
ELECTROPHORESIS, 2005, 26 (24) :4664-4673
[3]  
Bruin GJM, 2000, ELECTROPHORESIS, V21, P3931, DOI 10.1002/1522-2683(200012)21:18<3931::AID-ELPS3931>3.0.CO
[4]  
2-M
[5]   Determination of vitamins in food based on supercritical fluid extraction prior to micellar electrokinetic capillary chromatographic analyses of individual vitamins [J].
Buskov, S ;
Moller, P ;
Sorensen, H ;
Sorensen, JC ;
Sorensen, S .
JOURNAL OF CHROMATOGRAPHY A, 1998, 802 (01) :233-241
[6]  
CEVILLEN AG, 2006, ANAL CHIM ACTA, V562, P137
[7]   Fabrication and performance of a three-dimensionally adjustable device for the amperometric detection of microchip capillary electrophoresis [J].
Chen, G ;
Bao, HM ;
Yang, PY .
ELECTROPHORESIS, 2005, 26 (24) :4632-4640
[8]   Recent advances in the application of capillary electromigration methods for food analysis [J].
Cifuentes, A .
ELECTROPHORESIS, 2006, 27 (01) :283-303
[9]   Clinical potential of microchip capillary electrophoresis systems [J].
Colyer, CL ;
Tang, T ;
Chiem, N ;
Harrison, DJ .
ELECTROPHORESIS, 1997, 18 (10) :1733-1741
[10]  
Dolník V, 2000, ELECTROPHORESIS, V21, P41, DOI 10.1002/(SICI)1522-2683(20000101)21:1<41::AID-ELPS41>3.0.CO