Oxygen saturation measurements in canine blood containing hemoglobin glutamer-200 (bovine): In vitro validation of the NOVA CO-Oximeter

被引:11
作者
Jahr, JS
Driessen, B
Lurie, F
Tang, ZP
Louie, RF
Kost, G
机构
[1] Univ Penn, New Bolton Ctr, Sch Vet Med, Dept Clin Studies, Kennett Sq, PA 19348 USA
[2] Univ Calif Davis, Med Ctr, Sch Med, Dept Anesthesiol & Pain Med, Sacramento, CA 95817 USA
[3] Univ Calif Davis, Med Ctr, Sch Med, Dept Med Pathol, Sacramento, CA 95817 USA
关键词
hemoglobin-based oxygen carrier; hemoglobin glutamer-200 (bovine); NOVA CO-Oximeter; oxygen content; oxygen saturation; tonometry;
D O I
10.1111/j.1939-165X.2001.tb00255.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
This study was designed to validate in vitro oxygen saturation (SO2) measurements with the NOVA CO-Oximeter (Nova Biomedical Corp, Waltham, Mass, USA) in canine blood containing hemoglobin (Hb) glutamer-200 bovine (Hb-200; Oxyglobin, Biopure, Cambridge, Mass, USA) as a Hb-based oxygen carrier recently introduced into clinical practice. In the first set of experiments, stored blood from 6 mixed-breed canine blood donors was used. Target Po-2 levels were reached in aliquots of blood samples by tonometry. Oxygen saturation was then measured with the test device and calculated based on known Po-2 values. In the second set of experiments, total oxygen content was directly measured by means of an oxygen-specific electrode in aliquots of fresh whole arterial, venous, and mixed (arterial-venous) blood withdrawn from the same canine blood donors. Hb-200 was added to those blood samples to yield plasma Hb concentrations of 1.62, 3.25, 6.50, and 9.75 g/dL. Based on Hb content and SO2 measured by the NOVA CO-Oximeter in these samples, total oxygen content was also calculated for each sample and compared with measured values. A strong correlation was found between SO2 values measured with the co-oximeter in samples after tonometry, and calculated SO2 based on known Po-2. Directly measured total blood O-2 content varied by less than or equal to5% from values computed based on co-oximeter measurements of Hb content and SO2. These results did not change with different levels of oxygenation of the samples or different plasma Hb-200 concentrations. In conclusion, the NOVA CO-Oximeter is an accurate analyzer for measurement of SO2 after Hb-200 administration to canine blood. (C) 2001 American Society for Veterinary Clinical Pathology.
引用
收藏
页码:39 / 45
页数:7
相关论文
共 34 条
[1]   MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES [J].
ALTMAN, DG ;
BLAND, JM .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) :307-317
[2]  
[Anonymous], APPL RESP PHYSL
[3]  
Braden TD, 1999, AM J VET RES, V60, P1337
[4]   Fluid therapy and newer blood products [J].
Broadstone, RV .
VETERINARY CLINICS OF NORTH AMERICA-SMALL ANIMAL PRACTICE, 1999, 29 (03) :611-+
[5]  
Callas DD, 1997, CLIN CHEM, V43, P1744
[6]   Blood substitutes: Fluids, drugs, or miracle solutions? [J].
Dietz, NM ;
Joyner, MJ ;
Warner, MA .
ANESTHESIA AND ANALGESIA, 1996, 82 (02) :390-405
[7]   OXYGEN BINDING-CAPACITY OF HUMAN HEMOGLOBIN - HUFNERS FACTOR REDETERMINED [J].
DIJKHUIZEN, P ;
BUURSMA, A ;
FONGERS, TME ;
GERDING, AM ;
OESEBURG, B ;
ZIJLSTRA, WG .
PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1977, 369 (03) :223-231
[8]  
GILLESPIE JR, 1990, J AM VET MED ASSOC, V156, P717
[9]  
Harvey J. W., 1997, Clinical biochemistry of domestic animals., P157, DOI 10.1016/B978-012396305-5/50008-7
[10]  
HASKINS SC, 1998, P 22 ANN WALT OSU S, P7