Use of radial arterial tonometric continuous blood pressure measurement in cardiovascular reactivity studies

被引:24
作者
Nelesen, RA [1 ]
Dimsdale, JE [1 ]
机构
[1] Univ Calif San Diego, Dept Psychiat, La Jolla, CA 92093 USA
关键词
arterial tonometric; continuous blood pressure; non-invasive blood pressure; cardiovascular reactivity;
D O I
10.1097/00126097-200210000-00002
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background Non-invasive continuous blood pressure monitoring has, for the past decade, been performed using the Finapres monitor. Radial tonometry is a new methodology for measuring non-invasive beat-to-beat blood pressure. This technique has been evaluated for other applications, but its use in cardiovascular reactivity testing has not yet been reported. Design A convenience sample was used to compare a radial tonometric device (Colin Pilot) with intermittent brachial blood pressure measurement (Critikon Dinamap) during reactivity testing. A second comparison was made between an Ohmeda Finapres and a Dinamap. Methods Twenty subject-testing sessions during a reactivity protocol were evaluated for blood pressure determination. Separate groups of subjects were used for Pilot versus Dinamap comparisons and Finapres versus Dinamap comparisons. Blood pressure signals from 15 subjects with Pilot and 15 subjects with Finapres data were evaluated for artifacts. Signals were also compared in a subject with Raynaud's phenomenon. Results Both devices tracked blood pressure well. The Finapres significantly overestimated systolic blood pressure (P < 0.001). The correlations between the Pilot and Dinamap were consistently higher than those between the Finapres and Dinamap. There were more artifacts (3.2%) With the Finapres than with the Pilot (0.2%). In the subject with Raynaud's phenomenon, the Finapres overestimated blood pressure by more than 40 mmHg and showed more than 75% of artifacts. On the other hand, the Pilot tracked blood pressure in the same subject without error. Conclusions Radial tonometric continuous beat-to-beat blood pressure monitoring with the Colin Pilot was associated with a low artifact rating and high accuracy of blood pressure measurement during reactivity testing. (C) 2002 Lippincott Williams Wilkins.
引用
收藏
页码:259 / 263
页数:5
相关论文
共 16 条
[1]   Effects of age and gender on autonomic control of blood pressure dynamics [J].
Barnett, SR ;
Morin, RJ ;
Kiely, DK ;
Gagnon, M ;
Azhar, G ;
Knight, EL ;
Nelson, JC ;
Lipsitz, LA .
HYPERTENSION, 1999, 33 (05) :1195-1200
[2]   Continuous blood pressure determination during exercise ECG recording [J].
Barschdorff, D ;
Erig, M .
BIOMEDIZINISCHE TECHNIK, 1998, 43 (03) :34-39
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Spectral analysis of blood pressure: is it pointing in a useful direction? [J].
Cameron, JD ;
McGrath, BP .
CLINICAL SCIENCE, 1999, 97 (02) :253-254
[5]   On the reliable assessment of cardiovascular recovery: An application of curve-fitting techniques [J].
Christenfeld, N ;
Glynn, LM ;
Gerin, W .
PSYCHOPHYSIOLOGY, 2000, 37 (04) :543-550
[6]  
HAN K, 1991, BIOELECTRIC PHENOMEN, P660
[7]   Fifteen years experience with finger arterial pressure monitoring: assessment of the technology [J].
Imholz, BPM ;
Wieling, W ;
van Montfrans, GA ;
Wesseling, KH .
CARDIOVASCULAR RESEARCH, 1998, 38 (03) :605-616
[8]  
Imholz BPM, 1996, Z KARDIOL, V85, P76
[9]   Autonomic cardiovascular regulation in obesity [J].
Laederach-Hofmann, K ;
Mussgay, L ;
Rúddel, H .
JOURNAL OF ENDOCRINOLOGY, 2000, 164 (01) :59-66
[10]   CARDIOVASCULAR REACTIVITY TO A NEW MENTAL STRESS TEST - THE MAZE TEST [J].
MOUNIERVEHIER, C ;
GIRARD, A ;
CONSOLI, S ;
LAUDE, D ;
VACHERON, A ;
ELGHOZI, JL .
CLINICAL AUTONOMIC RESEARCH, 1995, 5 (03) :145-150