Textile analysis of heavy weight, mid-weight, and light weight polypropylene mesh in a porcine ventral hernia model

被引:144
作者
Cobb, William S. [1 ]
Burns, Justin M. [1 ]
Peindl, Richard D. [1 ]
Carbonell, Alfredo M. [1 ]
Matthews, Brent D. [1 ]
Kercher, Kent W. [1 ]
Heniford, B. Todd [1 ]
机构
[1] Carolinas Med Ctr, Carolinas Laparascop & Adv Surg Program, Dept Gen Surg, Charlotte, NC 28203 USA
关键词
surgical mesh; polypropylenes; biocompatible materials; materials testing; hernia; ventral; foreign-body reaction;
D O I
10.1016/j.jss.2006.05.022
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background. The purpose of this study was to assess the burst strength and stiffness of heavy weight (HW), mid-weight (MW), and light weight (LW) polypropylene mesh pre-implantation and 5 months post-implantation in a porcine ventral hernia model. Materials and methods. HM (95 g/cm(2)), MW (45 g/cm(2)), and LW (28 g/cm(2)) polypropylene were tested as dry samples (n = 8/mesh) before implantation using a stamp strain machine. Three weeks after creating four hernia defects in each mini-pig (n = 18), the polypropylene meshes (8 x 10 cm; 24/mesh type) were implanted in the preperitoneal space, ensuring 2 cm overlap circumferentially. At 5 months, the mesh was harvested en bloc with the abdominal wall. Testing of burst strength and stiffness was repeated on the ex vivo specimens. Results. After implantation in the pig for 5 months, the mean burst strength and stiffness of HW and MW meshes remains the same. LW mesh with an absorbable monofilament in its weave has significant reductions in mean burst strength (P = 0.01) and mean stiffness (P < 0.0001). The burst strength for all meshes tested was much higher than measured for the abdominal wall fascia alone. The degree of shrinkage of the various weight meshes did not differ. Conclusion. A reduction in mean burst strength and stiffness occurs after 5 months implantation of LW polypropylene mesh with an absorbable monofilament. All meshes exhibited burst strengths that were much greater than the burst strength of the abdominal wall fascia alone. After tissue incorporation, the LW polypropylene mesh maintains mean burst strength comparable to MW polypropylene mesh, while becoming less stiff than HW mesh. Long-term, this may contribute to more physiological abdominal wall compliance after LW polypropylene mesh implantation. (c) 2006 Elsevier Inc. All rights reserved.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 24 条
[1]  
Burger JWA, 2004, ANN SURG, V240, P578
[2]   Normal intraabdominal pressure in healthy adults [J].
Cobb, WS ;
Burns, JM ;
Kercher, KW ;
Matthews, BD ;
Norton, HJ ;
Heniford, BT .
JOURNAL OF SURGICAL RESEARCH, 2005, 129 (02) :231-235
[3]  
Greca F H, 2001, Hernia, V5, P59
[4]   Laparoscopic repair of ventral hernias nine years' experience with 850 consecutive hernias [J].
Heniford, BT ;
Park, A ;
Ramshaw, BJ ;
Voeller, G .
ANNALS OF SURGERY, 2003, 238 (03) :391-399
[5]  
HESSELINK VJ, 1993, SURG GYNECOL OBSTET, V176, P228
[6]  
Junge K, 2001, Hernia, V5, P113
[7]  
Klinge U, 1999, EUR J SURG, V165, P665
[8]  
Klinge U, 1998, EUR J SURG, V164, P965
[9]  
Klinge U, 1996, CHIRURG, V67, P229
[10]  
Klinge U, 1998, EUR J SURG, V164, P951