An implantable neuroprosthesis for standing and walking in paraplegia: 5-year patient follow-up

被引:79
作者
Guiraud, David [1 ]
Stieglitz, Thomas
Koch, Klaus Peter
Divoux, Jean-Louis
Rabischong, Pierre
机构
[1] LIRMM, INRIA, DEMAR Project, F-34392 Montpellier 5, France
[2] Fraunhofer Inst Biomed Engn, D-66386 Sankt Ingbert, Germany
[3] Labs MXM, F-06224 Vallauris, France
[4] Propara Ctr, F-34195 Montpellier 5, France
关键词
D O I
10.1088/1741-2560/3/4/003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present the results of a 5-year patient follow-up after implantation of an original neuroprosthesis. The system is able to stimulate both epimysial and neural electrodes in such a way that the complete flexor-extensor chain of the lower limb can be activated without using the withdrawal reflex. We demonstrate that standing and assisted walking are possible, and the results have remained stable for 5 years. Nevertheless, some problems were noted, particularly regarding the muscle response on the epimysial channels. Analysis of the electrical behaviour and thresholds indicated that the surgical phase is crucial because of the sensitivity of the functional responses to electrode placement. Neural stimulation proved to be more efficient and more stable over time. This mode requires less energy and provides more selective stimulation. This FES system can be improved to enable balanced standing and less fatiguing gait, but this will require feedback on event detection to trigger transitions between stimulation sequences, as well as feedback to the patient about the state of his lower limbs.
引用
收藏
页码:268 / 275
页数:8
相关论文
共 19 条
[1]  
Arabi K, 1999, IEEE Trans Rehabil Eng, V7, P204, DOI 10.1109/86.769411
[2]   Paraplegia: Prolonged closed-loop standing with implanted nucleus FES-22 stimulator and Andrews' foot-ankle orthosis [J].
Davis, R ;
Houdayer, T ;
Andrews, B ;
Emmons, S ;
Patrick, J .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1997, 69 (1-4) :281-287
[3]   Lumbar root stimulation for restoring leg function: Stimulator and measurement of muscle actions [J].
Donaldson, NDN ;
Perkins, TA ;
Worley, ACM .
ARTIFICIAL ORGANS, 1997, 21 (03) :247-249
[4]  
El Makssoud H, 2004, IEEE INT CONF ROBOT, P1282
[5]  
GUIRAUD D, 2001, P CDROM IFESS 6 C CL
[6]  
GUIRAUD D, 2000, P IFESS 5 C AALB DEN, P292
[7]  
Kobetic R, 1999, IEEE Trans Rehabil Eng, V7, P390, DOI 10.1109/86.808942
[8]  
Kobetic R, 1997, IEEE Trans Rehabil Eng, V5, P23, DOI 10.1109/86.559346
[9]  
KOCH KP, 2000, IFESS 5 C AALB DENM, P261
[10]   BION™ system for distributed neural prosthetic interfaces [J].
Loeb, GE ;
Peck, RA ;
Moore, WH ;
Hood, K .
MEDICAL ENGINEERING & PHYSICS, 2001, 23 (01) :9-18