Stereotactic localization with magnetic resonance imaging: A phantom study to compare the accuracy obtained using two-dimensional and three-dimensional data acquisitions

被引:59
作者
Walton, L [1 ]
Hampshire, A [1 ]
Forster, DMC [1 ]
Kemeny, AA [1 ]
机构
[1] ROYAL HALLAMSHIRE HOSP,NATL CTR STEREOTAC RADIOSURG,SHEFFIELD S10 2JF,S YORKSHIRE,ENGLAND
关键词
magnetic resonance distortion; radiosurgery; stereotactic localization; stereotaxy;
D O I
10.1097/00006123-199707000-00027
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: To compare the accuracy of stereotactic localization using magnetic resonance imaging (Siemens 1.5-T Magnetom; Siemens, Erlangen, Germany) with two-dimensional and three-dimensional data acquisition techniques. METHODS: A phantom study was performed in which the coordinates of an array of rods were determined from images in both two-dimensional and three-dimensional studies and compared with measured values in a series of transverse, coronal, and sagittal images. RESULTS: The results demonstrated a distinct advantage in using three-dimensional acquisition; an error greater than 2 mm was identified in only 0.8% of the imaged volume, compared with 12% of the imaged volume in the two-dimensional study. CONCLUSION: The results indicate that more accurate stereotactic localization is achieved with a three-dimensional acquisition.
引用
收藏
页码:131 / 137
页数:7
相关论文
共 20 条
[1]   3-DIMENSIONAL MR-IMAGING AND DISPLAY OF INTRACRANIAL DISEASE - IMPROVEMENTS WITH THE MP-RAGE SEQUENCE AND GADOLINIUM [J].
BRANTZAWADZKI, MN ;
GILLAN, GD ;
ATKINSON, DJ ;
EDALATPOUR, N ;
JENSEN, M .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (04) :656-662
[2]  
ELSTER AD, 1993, RADIOLOGY, V186, P1
[3]   TREATMENT VOLUME SHAPING WITH SELECTIVE BEAM BLOCKING USING THE LEKSELL GAMMA-UNIT [J].
FLICKINGER, JC ;
MAITZ, A ;
KALEND, A ;
LUNSFORD, LD ;
WU, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 19 (03) :783-789
[4]   TREATMENT PLANNING FOR GAMMA KNIFE RADIOSURGERY WITH MULTIPLE ISOCENTERS [J].
FLICKINGER, JC ;
LUNSFORD, LD ;
WU, A ;
MAITZ, AH ;
KALEND, AM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (06) :1495-1501
[5]   A 3-DIMENSIONAL VOLUME VISUALIZATION PACKAGE APPLIED TO STEREOTAXIC RADIOSURGERY TREATMENT PLANNING [J].
GEHRING, MA ;
MACKIE, TR ;
KUBSAD, SS ;
PALIWAL, BR ;
MEHTA, MP ;
KINSELLA, TJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (02) :491-500
[6]   DIAGNOSTIC-IMAGING AS A MEASURING DEVICE FOR STEREOTAXIC NEUROSURGERY [J].
JONES, AP .
PHYSIOLOGICAL MEASUREMENT, 1993, 14 (02) :91-112
[7]   A COMPARISON BETWEEN MAGNETIC-RESONANCE-IMAGING AND COMPUTED-TOMOGRAPHY FOR STEREOTAXIC COORDINATE DETERMINATION [J].
KONDZIOLKA, D ;
DEMPSEY, PK ;
LUNSFORD, D ;
HEILBRUN, MP ;
KESTLE, JRW ;
DOLAN, EJ ;
KANAL, E ;
TASKER, RR .
NEUROSURGERY, 1992, 30 (03) :402-407
[8]   IMAGE FUSION FOR STEREOTAXIC RADIOTHERAPY AND RADIOSURGERY TREATMENT PLANNING [J].
KOOY, HM ;
VANHERK, M ;
BARNES, PD ;
ALEXANDER, E ;
DUNBAR, SF ;
TARBELL, NJ ;
MULKERN, RV ;
HOLUPKA, EJ ;
LOEFFLER, JS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (05) :1229-1234
[9]   TREATMENT PLANNING FOR STEREOTAXIC RADIOSURGERY OF INTRACRANIAL LESIONS [J].
KOOY, HM ;
NEDZI, LA ;
LOEFFLER, JS ;
ALEXANDER, E ;
CHENG, CW ;
MANNARINO, EG ;
HOLUPKA, EJ ;
SIDDON, RL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03) :683-693
[10]   IMAGE DIRECTED STEREOTAXIC SURGERY FOR BRAIN-STEM LESIONS [J].
KRATIMENOS, GP ;
NOUBY, RM ;
BRADFORD, R ;
PELL, MF ;
THOMAS, DGT .
ACTA NEUROCHIRURGICA, 1992, 116 (2-4) :164-170