Quantitative determination by using HPLC and GLC methods for cocaine HCl in synthetic binary mixtures with procaine HCl, lidocaine HCl and caffeine

被引:13
作者
Atay, O [1 ]
Oztop, F [1 ]
机构
[1] NARCOT LAB GENDARMARIE GEN COMMAND, ANKARA, TURKEY
关键词
cocaine HCl; procaine HCl; lidocaine HCl; caffeine; HPLC; GLC; binary mixtures; quantitative determination; PERFORMANCE LIQUID-CHROMATOGRAPHY; ILLICIT COCAINE; DERIVATIVE SPECTROPHOTOMETRY; GAS-CHROMATOGRAPHY; LOCAL-ANESTHETICS; SAMPLES; DRUGS; IDENTIFICATION; HEROIN; ABUSE;
D O I
10.1080/00032719708001802
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Cocaine HCl is a substance which creates psychological and physical dependence. Usually it is available on the market being diluted by other substances, like local anesthetics, analeptics and CNS stimulating agents and inert substance. In this study, HPLC and GLC methods were applied for qualitative and quantitative determinations of synthetic binary mixtures. In the HPLC determination, mu Bondopack C-18 10 mu m. column system, a mobile phase consisting of methanol-water-phosphoric acid-1% hexylamine (75:175:250:3,5) and U.V. detection by photodiodearray (196-500 nm) were used. The linear concentration areas were found in a range of 2.5-25 mu g/mL. The R.S.D percentages for cocaine HCl, procaine HCl, lidocaine HCl and caffeine were found as 0.922, 0.568, 1.18 and 1.04, respectively. In the GLC determination, two different column systems, a 2% OV-17- Gas Chrom W-HP 100-200 mesh filled column and a 0.25 SE-52 fused silica capillary column, were used. Nitrogen was used in a filled column and helium was used in a capillary column. Mobile phase flow rates were set as 30 mL/min and a flame ionization detector was used with both column systems. The linear concentration intervals were found in a range of 2-25 mu L/mL in both methods. The R.S.D. for cocaine HCl, procaine HCl, lidocaine HCl, and caffeine were found to be 0.907, 0.948, 0.770, 0.901 in the filled column. For cocaine HCl, procaine HCl and caffeine R.S.D.'s of 0.774, 0.809, 0.814 were found, when a capillary column was used. In quantitative determinations, antipyrine was chosen as internal standard in the HPLC and GLC methods.
引用
收藏
页码:565 / 584
页数:20
相关论文
共 29 条
[1]   IDENTIFICATION OF COCAINE IN COCAINE LIDOCAINE MIXTURES (ROCK COCAINE) AND OTHER ILLICIT COCAINE PREPARATIONS USING DERIVATIVE ABSORPTION-SPECTROSCOPY [J].
ARUFEMARTINEZ, MI ;
ROMEROPALANCO, JL .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1988, 12 (04) :192-196
[2]  
ARUFEMARTINEZ MI, 1989, J ANAL TOXICOL, V13, P337, DOI 10.1093/jat/13.6.337
[3]  
ARUFEMARTINEZ MI, 1989, J ANAL TOXICOL, V13, P345
[4]  
BARNICOMPARINI I, 1983, J CHROMATOGR, V279, P609
[5]  
BEYER HK, 1987, DTSCH APOTH ZTG, V127, P2037
[6]   A CAPILLARY COLUMN GAS-CHROMATOGRAPHIC METHOD FOR THE IDENTIFICATION OF DRUGS OF ABUSE IN URINE SAMPLES [J].
CALDWELL, R ;
CHALLENGER, H .
ANNALS OF CLINICAL BIOCHEMISTRY, 1989, 26 :430-443
[7]   A QUANTITATIVE DENSITOMETRIC DETERMINATION OF HEROIN AND COCAINE SAMPLES BY HIGH-PERFORMANCE THIN-LAYER CHROMATOGRAPHY [J].
DELLACASA, E ;
MARTONE, G .
FORENSIC SCIENCE INTERNATIONAL, 1986, 32 (02) :117-120
[8]  
DRUMGOLD MR, 1990, LC GC-MAG SEP SCI, V8, P392
[9]  
FERNANDEZABEDUL TM, 1991, ELECTROANAL, V3, P409
[10]  
FUNK W, 1991, Journal of Planar Chromatography - Modern TLC, V4, P123