Blood microdialysis in humans: A new method for monitoring plasma compounds

被引:20
作者
Paez, X
Hernandez, L
机构
[1] Laboratory of Behavioral Physiology, Universidad de Los Andes, Mérida 5101A
[2] Mérida 5101A
关键词
venous microdialysis probe; humans; chemical plasma monitoring; glucose plasma levels; unbound substances;
D O I
10.1016/S0024-3205(97)00586-9
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
A new venous microdialysis probe for humans was developed. The active area was made with cellulose membrane (220 mm outside diameter, length 20 mm, and 6000 molecular weight cut off). Glucose measurements were used to test it. The relative recovery was 50% in dialysates obtained in vitro, the step response was 91.8% in the first 2 min. The in vivo recovery was 65.8% obtained by the no net flux method, with the probe placed in the cubital vein of the subjects arms. Corrected glucose in dialysates was non significantly different from plasma glucose simultaneously collected. In healthy male volunteers from 18 to 37 years of age, mean 22.1 years, the glucose levels in the blood dialysates were measured every 30 min, 60 min before and 150 min after an oral load of 75 g of glucose or vehicle. Glucose increased significantly after the load. The differences with the control group were significant at 60 and 90 min after the treatment. The results showed the easy and safe application of this technique to monitor endogenous and exogenous compounds in the extracellular compartment of blood in out patients.
引用
收藏
页码:847 / 856
页数:10
相关论文
共 32 条
[1]  
ANDERSON C, 1995, IRRITANT DERMATITIS, P121
[2]   MICRODIALYSIS OF ADIPOSE-TISSUE AND BLOOD FOR INVIVO LIPOLYSIS STUDIES [J].
ARNER, P ;
BOLINDER, J ;
ELIASSON, A ;
LUNDIN, A ;
UNGERSTEDT, U .
AMERICAN JOURNAL OF PHYSIOLOGY, 1988, 255 (05) :E737-E742
[3]   ESTIMATION OF RENAL INTERSTITIAL ADENOSINE AND PURINE METABOLITES BY MICRODIALYSIS [J].
BARANOWSKI, RL ;
WESTENFELDER, C .
AMERICAN JOURNAL OF PHYSIOLOGY, 1994, 267 (01) :F174-F182
[4]   LONG-TERM CONTINUOUS GLUCOSE MONITORING WITH MICRODIALYSIS IN AMBULATORY INSULIN-DEPENDENT DIABETIC-PATIENTS [J].
BOLINDER, J ;
UNGERSTEDT, U ;
ARNER, P .
LANCET, 1993, 342 (8879) :1080-1085
[5]  
CHEN ZD, 1993, J PHARMACOL TOXICOL, V29, P111
[6]   MICRODIALYSIS PROBE FOR TRANSCUTANEOUS MONITORING OF ETHANOL AND GLUCOSE IN HUMANS [J].
DEBOER, J ;
PLIJTERGROENDIJK, H ;
KORF, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1993, 75 (06) :2825-2830
[7]  
DELEU D, 1995, J PHARMACOL EXP THER, V273, P1323
[8]  
EKSTROM PO, 1995, CANCER CHEMOTH PHARM, V36, P283, DOI 10.1007/s002800050324
[9]   INCREASE IN SEROTONIN LEVELS IN THE DOG ILEUM AND BLOOD BY CISPLATIN AS MEASURED BY MICRODIALYSIS [J].
FUKUI, H ;
YAMAMOTO, M ;
ANDO, T ;
SASAKI, S ;
SATO, S .
NEUROPHARMACOLOGY, 1993, 32 (10) :959-968
[10]   MICRODIALYSIS IN CUTANEOUS PHARMACOLOGY - KINETIC-ANALYSIS OF TRANSDERMALLY DELIVERED NICOTINE [J].
HEGEMANN, L ;
FORSTINGER, C ;
PARTSCH, B ;
LAGLER, I ;
KROTZ, S ;
WOLFF, K .
JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1995, 104 (05) :839-843