Accurate characterization of Monte Carlo calculated electron beams for radiotherapy

被引:122
作者
Ma, CM
Faddegon, BA
Rogers, DWO
Mackie, TR
机构
[1] TORONTO SUNNYBROOK REG CANC CTR, TORONTO, ON, CANADA
[2] UNIV WISCONSIN, DEPT MED PHYS, MADISON, WI 53706 USA
关键词
radiotherapy; accelerator simulation; beam characterization; EGS4 Monte Carlo; beam model;
D O I
10.1118/1.597908
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Monte Carlo studies of dose distributions in patients treated with radiotherapy electron beams would benefit from generalized models of clinical beams if such models introduce little error into the dose calculations. Methodology is presented for the design of beam models, including their evaluation in terms of how well they preserve the character of the clinical beam, and the effect of the beam models on the accuracy of dose distributions calculated with Monte Carlo. This methodology has been used to design beam models for electron beams from two linear accelerators, with either a scanned beam or a scattered beam. Monte Carlo simulations of the accelerator heads are done in which a record is kept of the particle phase-space, including the charge, energy, direction, and position of every particle that emerges from the treatment head, along with a tag regarding the details of the particle history. The character of the simulated beams are studied in detail and used to design various beam models from a simple point source to a sophisticated multiple-source model which treats particles from different parts of a linear accelerator as from different sub-sources. Dose distributions calculated using both the phase-space data and the multiple-source model agree within 2%, demonstrating that the model is adequate for the purpose of Monte Carlo treatment planning for the beams studied. Benefits of the beam models over phase-space data for dose calculation are shown to include shorter computation time in the treatment head simulation and a smaller disk space requirement, both of which impact on the clinical utility of Monte Carlo treatment planning. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:401 / 416
页数:16
相关论文
共 29 条
[1]   MONTE-CARLO TECHNIQUES IN MEDICAL RADIATION PHYSICS [J].
ANDREO, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (07) :861-920
[2]   STOPPING-POWER RATIOS AND THEIR UNCERTAINTIES FOR CLINICAL ELECTRON-BEAM DOSIMETRY [J].
ANDREO, P ;
FRANSSON, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (12) :1847-1861
[3]  
Bielajew A., 1990, The Dosimetry of Ionizing Radiation, Volume, P427
[4]  
BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
[5]  
BIELAJEW AF, 1987, USE COMPUTERS RAD TH, P65
[6]  
BRUN R, 1992, CERN COMPUTER CTR
[7]   ELECTRON DOSE DISTRIBUTIONS IN EXPERIMENTAL PHANTOMS - A COMPARISON WITH 2D PENCIL BEAM CALCULATIONS [J].
CYGLER, J ;
BATTISTA, JJ ;
SCRIMGER, JW ;
MAH, E ;
ANTOLAK, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (09) :1073-1086
[8]  
DING GX, 1995, PIRS0439 NRC NAT RES
[9]   A MODEL FOR ELECTRON-BEAM APPLICATOR SCATTER [J].
EBERT, MA ;
HOBAN, PW .
MEDICAL PHYSICS, 1995, 22 (09) :1419-1429
[10]   ELECTRON-BEAM DOSE CALCULATIONS [J].
HOGSTROM, KR ;
MILLS, MD ;
ALMOND, PR .
PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (03) :445-459