Monte Carlo simulations of the differential beam hardening effect of a flattening filter on a therapeutic x-ray beam

被引:39
作者
Lee, PC
机构
[1] Loyola University, Loyola-Hines Dept. of Radiotherapy, Maywood
关键词
Monte Carlo simulation; flattening filter; beam hardening effect; therapeutic x-ray beam;
D O I
10.1118/1.598037
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A quantitative study of the differential beam hardening effect of the flattening filter on the 6-MV beam of Clinac 2100C has been conducted with Monte Carlo simulations using EGS4 code. The fluence-weighted photon energy of the unfiltered beam decreases from 1.35 MeV at central axis (CAX) to 1.22 MeV at an off-axis distance (OAD) of 20.0 cm. Compared to the unfiltered beam, the fluence-weighted photon energy of the filtered beam increases to 1.93 MeV at CAX and to 1.36 MeV at an OAD of 20.0 cm, respectively. The beam hardening effect was found to decrease with increasing OAD. The energy fluence of the unfiltered beam was found to be 2.1 times higher at CAX than at an OAD of 20.0 cm. With the differential filtration of the flattening filter, the photon energy fluence reduced to 44% and 78% at CAX and an OAD of 20.0 cm, respectively, resulting in the energy fluence of the filtered beam being Bat from CAX to an OAD of 20.0 cm. The differential transmission ratios between the high energy and low energy photons decrease as the OAD increases. The percentage depth doses (PDDs) at field size of 10.0 cmx10.0 cm for both the filtered and unfiltered 6-MV beams at CAX and at an OAD of 15.0 cm were calculated with a Monte Carlo technique based on the simulated spectra and fluence. The calculated PDDs were found to be consistent with the measured data for the filtered beam at CAX and an OAD of 15.0 cm. The beam quality (BQ) of the filtered beam was found to be higher than that of the unfiltered beam at the same OAD. The BQ of the filtered beam at CAX is also higher than that of the same beam at an OAD of 15.0 cm. All the above results quantitatively demonstrate the differential beam hardening effects of a flattening filter-on a therapeutic x-ray beam. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:1485 / 1489
页数:5
相关论文
共 10 条
[1]   ANALYTIC MODELING OF PHOTON SCATTER FROM FLATTENING FILTERS IN PHOTON THERAPY BEAMS [J].
AHNESJO, A .
MEDICAL PHYSICS, 1994, 21 (08) :1227-1235
[2]  
BIELAJEW AF, 1989, PIRS0203 NAT RES COU
[3]   RADIATION BEAM CHARACTERISTICS OF A 22 MEV MICROTRON [J].
BRAHME, A ;
SVENSSON, H .
ACTA RADIOLOGICA ONCOLOGY, 1979, 18 (03) :244-272
[4]  
HAN K, 1995, MED PHYS
[5]  
KUBAS SS, 1990, INT J RADIAT ONCOL, V19, P1027
[6]   THE QUALITY OF HIGH-ENERGY X-RAY-BEAMS [J].
LARIVIERE, PD .
BRITISH JOURNAL OF RADIOLOGY, 1989, 62 (737) :473-481
[7]   MONTE-CARLO MODEL OF PHOTON BEAMS USED IN RADIATION-THERAPY [J].
LOVELOCK, DMJ ;
CHUI, CS ;
MOHAN, R .
MEDICAL PHYSICS, 1995, 22 (09) :1387-1394
[8]  
MCCALL R C, 1978, Medical Physics (Woodbury), V5, P518, DOI 10.1118/1.594487
[9]   ENERGY AND ANGULAR-DISTRIBUTIONS OF PHOTONS FROM MEDICAL LINEAR ACCELERATORS [J].
MOHAN, R ;
CHUI, C ;
LIDOFSKY, L .
MEDICAL PHYSICS, 1985, 12 (05) :592-597
[10]  
NELSON WR, 1985, SLACREPORT265 STANF