Noncontact atomic force microscopy simulator with phase-locked-loop controlled frequency detection and excitation

被引:35
作者
Nony, Laurent
Baratoff, Alexis
机构
[1] Univ Aix Marseille 3, UMR CNRS 6137, L2MP, F-13397 Marseille 20, France
[2] Univ Basel, NCCR Nanoscale Sci, CH-4056 Basel, Switzerland
关键词
D O I
10.1103/PhysRevB.74.235439
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A simulation of an atomic force microscope operating in the constant amplitude dynamic mode is described. The implementation mimics the electronics of a real setup including a digital phase-locked loop (PLL). The PLL is not only used as a very sensitive frequency detector, but also to generate the time-dependent phase-shifted signal driving the cantilever. The optimum adjustments of individual functional blocks and their joint performance in typical experiments are determined in detail. Prior to testing the complete setup, the performances of the numerical PLL and of the amplitude controller were ascertained to be satisfactory compared to those of the real components. Attention is also focused on the issue of apparent dissipation, that is, of spurious variations in the driving amplitude caused by the nonlinear interaction occurring between the tip and the surface and by the finite response times of the various controllers. To do so, an estimate of the minimum dissipated energy that is detectable by the instrument upon operating conditions is given. This allows us to discuss the relevance of apparent dissipation that can be conditionally generated with the simulator in comparison to values reported experimentally. The analysis emphasizes that apparent dissipation can contribute to the measured dissipation up to 15% of the intrinsic dissipated energy of the cantilever interacting with the surface, but can be made negligible when properly adjusting the controllers, the PLL gains and the scan speed. It is inferred that the experimental values of dissipation usually reported in the literature cannot only originate in apparent dissipation, which favors the hypothesis of "physical" channels of dissipation.
引用
收藏
页数:19
相关论文
共 45 条
[1]   FREQUENCY-MODULATION DETECTION USING HIGH-Q CANTILEVERS FOR ENHANCED FORCE MICROSCOPE SENSITIVITY [J].
ALBRECHT, TR ;
GRUTTER, P ;
HORNE, D ;
RUGAR, D .
JOURNAL OF APPLIED PHYSICS, 1991, 69 (02) :668-673
[2]   Atomically resolved edges and kinks of NaCl islands on Cu(111):: Experiment and theory [J].
Bennewitz, R ;
Foster, AS ;
Kantorovich, LN ;
Bammerlin, M ;
Loppacher, C ;
Schär, S ;
Guggisberg, M ;
Meyer, E ;
Shluger, AL .
PHYSICAL REVIEW B, 2000, 62 (03) :2074-2084
[3]  
BEST RE, 1999, PHASE LOCKED LOOPS D
[4]   Analysis of mechanisms inducing damping in dynamic force microscopy:: surface viscoelastic behavior and stochastic resonance process [J].
Boisgard, R ;
Aimé, JP ;
Couturier, G .
APPLIED SURFACE SCIENCE, 2002, 188 (3-4) :363-371
[5]   Surface mechanical instabilities and dissipation under the action of an oscillating tip [J].
Boisgard, R ;
Aimé, JP ;
Couturier, G .
SURFACE SCIENCE, 2002, 511 (1-3) :171-182
[6]   Damping and instability in non-contact atomic force microscopy:: the contribution of the instrument [J].
Couturier, G ;
Boisgard, R ;
Dietzel, D ;
Aimé, JP .
NANOTECHNOLOGY, 2005, 16 (08) :1346-1353
[7]   Stability of an oscillating tip in noncontact atomic force microscopy:: Theoretical and numerical investigations [J].
Couturier, G ;
Nony, L ;
Boisgard, R ;
Aimé, JP .
JOURNAL OF APPLIED PHYSICS, 2002, 91 (04) :2537-2543
[8]   Noncontact atomic force microscopy:: Stability criterion and dynamical responses of the shift of frequency and damping signal [J].
Couturier, G ;
Boisgard, R ;
Nony, L ;
Aimé, JP .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (05) :2726-2734
[9]  
COUTURIER G, 2004, CONTRIBUTION INSTRUM
[10]  
Durig U, 1997, J APPL PHYS, V82, P3641, DOI 10.1063/1.365726