Electrode materials for nitric oxide detection

被引:22
作者
Allen, BW
Piantadosi, CA
Coury, LA
机构
[1] Duke Univ, Med Ctr, Dept Anesthesiol, Durham, NC 27710 USA
[2] Duke Univ, Med Ctr, Dept Med, Durham, NC 27710 USA
[3] Duke Univ, Dept Chem, Durham, NC 27708 USA
来源
NITRIC OXIDE-BIOLOGY AND CHEMISTRY | 2000年 / 4卷 / 01期
关键词
D O I
10.1006/niox.2000.0273
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Nitric oxide oxidation signals were compared for uniform test electrodes of platinum, iridium, palladium, rhodium, ruthenium, gold, graphite, and a nickel-porphyrin on graphite in deaerated phosphate-buffered saline (pH 7.0) at 35 degrees C. All tested materials detected NO. amperometrically, Current densities (A/M/cm(2) +/- S.D.) were Ir (0.021 +/- 0.002), Rh (0.088 +/- 0.012), graphite (0.117 +/- 0.018), Pd (0.118 +/- 0.033), Au (0.149 +/- 0.039), Pt (0.237 +/- 0.117), Ni (II)-tetra(3-methoxy-4-hydroxyphenyl) porphyrin on graphite (0.239 +/- 0.009), and Ru (0.680 +/- 0.058). NO. oxidation current on ruthenium was maximal at 675 mV (vs Ag/AgCl), nearly three times that on the next-best materials, platinum and Ni-porphyrin on graphite poised at 800 mV, The measured limit of detection for NO. on Ru was below 3 nM, Enhanced NO. oxidation current on ruthenium is apparently due to formation of nitrosyl-or chloronitrosyl-ruthenium complexes at the electrode surface. At fixed potentials above 675 mV, ruthenium exhibited an even larger NO. response, characterized by current flow opposite in polarity to an oxidation, which we hypothesize reflects suppression of the oxidative background current (presumably due to chloride oxidation or to the electrolysis of water) by a film consisting of nitrosyl- or chloronitrosyl-ruthenium complexes. The sensitive response of the ruthenium electrode to the direct oxidation of NO. may be useful in sensors for biomedical applications. (C) 2000 Academic Press.
引用
收藏
页码:75 / 84
页数:10
相关论文
共 22 条
  • [1] ALI H, 1997, ADV RHEUMATOLOGY, P1
  • [2] Fiber-optic nitric oxide-selective biosensors and nanosensors
    Barker, SLR
    Kopelman, R
    Meyer, TE
    Cusanovich, MA
    [J]. ANALYTICAL CHEMISTRY, 1998, 70 (05) : 971 - 976
  • [3] THE USE OF GOLD ELECTRODES IN THE ELECTROCHEMICAL DETECTION OF NITRIC-OXIDE IN AQUEOUS-SOLUTION
    BEDIOUI, F
    TREVIN, S
    DEVYNCK, J
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1994, 377 (1-2): : 295 - 298
  • [4] CONTINUOUS RECORDING OF BLOOD OXYGEN TENSIONS BY POLAROGRAPHY
    CLARK, LC
    WOLF, R
    GRANGER, D
    TAYLOR, Z
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1953, 6 (03) : 189 - 193
  • [5] EVANS MO, 1978, LANCET 1230, P1371
  • [6] GRIFFITH WP, 1967, CHEMISTRY RARER PLAT, P174
  • [7] HOARE JP, 1968, ELECTROCHEMISTRY OXY, P423
  • [8] In situ nitric oxide (NO) measurement by modified electrodes: NO labilized by photolysis of metal nitrosyl complexes
    Kudo, S
    Bourassa, JL
    Boggs, SE
    Sato, Y
    Ford, PC
    [J]. ANALYTICAL BIOCHEMISTRY, 1997, 247 (02) : 193 - 202
  • [9] LLOPIS JF, 1976, RUTHENIUM ENCY ELECT, V6
  • [10] NITRIC-OXIDE RELEASE FROM A SINGLE CELL MEASURED INSITU BY A PORPHYRINIC-BASED MICROSENSOR
    MALINSKI, T
    TAHA, Z
    [J]. NATURE, 1992, 358 (6388) : 676 - 678