Determination of forty two major and trace elements in USGS and NIST SRM glasses by laser ablation-inductively coupled plasma-mass spectrometry

被引:579
作者
Gao, S [1 ]
Liu, XM
Yuan, HL
Hattendorf, B
Günther, D
Chen, L
Hu, SH
机构
[1] NW Univ Xian, Key Lab Continental Dynam, Dept Geol, Xian 710069, Peoples R China
[2] China Univ Geosci, Fac Earth Sci, Wuhan 430074, Peoples R China
[3] ETH Honggerberg, Inorgan Chem Lab, CH-8093 Zurich, Switzerland
来源
GEOSTANDARDS NEWSLETTER-THE JOURNAL OF GEOSTANDARDS AND GEOANALYSIS | 2002年 / 26卷 / 02期
关键词
USGS glass reference materials; NIST SRM 610-614 trace elements; laser-ablation ICP-MS; homogeneity;
D O I
10.1111/j.1751-908X.2002.tb00886.x
中图分类号
P [天文学、地球科学];
学科分类号
07 ;
摘要
Forty two major (Na, Mg, Ti and Mn) and trace elements covering the mass range from Li to U in three USGS basalt glass reference materials BCR-2G, BHVO-2G and BIR-1G were determined using laser ablation-inductively coupled plasma-mass spectrometry. Calibration was performed using NIST SRM 610 in conjunction with internal standardisation using Ca. Determinations were also made on NIST SRM 612 and 614 as well as NIST SRM 610 as unknown samples, and included forty five major (Al and Na) and trace elements. Relative standard deviation (RSD) of determinations was below 10% for most elements in all the glasses under investigation. Consistent exceptions were Sri and Sb in BCR-2G, BHVO-2G and BIR-1G. For BCR-2G, BHVO-2G and BIR-1G, clear negative correlations on a logarithmic scale exist between RSD and concentration for elements lower than 1500 mug g(-1) with logarithmic correlation coefficients between -0.75 and -0.86. There is also a clear trend of increasing RSD with decreasing concentration from NIST SRM 610 through SRM 612 to SRM 614. These suggest that the difference in the scatter of apparent element concentrations is not due to chemical heterogeneity but reflects analytical uncertainty. It is concluded that all, these glasses are, overall, homogeneous on a scale of 60 mum. Our first results on BHVO-2G and BIR-1G showed that they generally agreed with BHVO-2/BHVO-1 and BIR-1 within 10% relative. Exceptions were Nb, To and Pb in BHVO-2G, which were 14-45% lower than reference values for BHVO-2 and BHVO-1. Be, Ni, Zn, Y, Zr, Nb, Sri, Sb, Gd, Tb, Er, Pb and U in BIR-1G were also exceptions. However, of these elements, Be, Nb, Sn, Sb, Gd, Tb, Pb and U gave results that were consistent within an uncertainty of 2s between our data and BIR-1 reference values. Results on NIST SRM 612 agreed well with published data, except for Mg and Sn. This was also true for elements with m/z greater than or equal to 85 (Rb) in the case of NIST SRM 614. The good agreement between measured and reference values for No and Mg in BCR-2G, BHVO-2G and BIR-1 G, and for A and Na in NIST SRM 610, 612 and 614 up to concentrations of at least several weight percent (which were possible to analyse due to the dynamic range of 108) indicates the suitability of this technique for major, minor and trace element determinations.
引用
收藏
页码:181 / 196
页数:16
相关论文
共 22 条
[1]   Laser ablation inductively coupled plasma mass spectrometry [J].
Alexander, ML ;
Smith, MR ;
Hartman, JS ;
Mendoza, A ;
Koppenaal, DW .
APPLIED SURFACE SCIENCE, 1998, 127 :255-261
[2]  
[Anonymous], 2001, MINERALOGICAL ASS CA
[3]   Determination of trace elements in geological samples by laser ablation inductively coupled plasma mass spectrometry [J].
Becker, JS ;
Dietze, HJ .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 365 (05) :429-434
[4]   Reference materials for geochemical studies: New analytical data by ICP-MS and critical discussion of reference values [J].
Dulski, P .
GEOSTANDARDS NEWSLETTER-THE JOURNAL OF GEOSTANDARDS AND GEOANALYSIS, 2001, 25 (01) :87-125
[5]   Laser ablation inductively coupled plasma mass spectrometry: achievements, problems, prospects [J].
Durrant, SF .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (09) :1385-1403
[6]   A simple method for the precise determination of >=40 trace elements in geological samples by ICPMS using enriched isotope internal standardisation [J].
Eggins, SM ;
Woodhead, JD ;
Kinsley, LPJ ;
Mortimer, GE ;
Sylvester, P ;
McCulloch, MT ;
Hergt, JM ;
Handler, MR .
CHEMICAL GEOLOGY, 1997, 134 (04) :311-326
[7]   Re-Os evidence for replacement of ancient mantle lithosphere beneath the North China craton [J].
Gao, S ;
Rudnick, RL ;
Carlson, RW ;
McDonough, WF ;
Liu, YS .
EARTH AND PLANETARY SCIENCE LETTERS, 2002, 198 (3-4) :307-322
[8]   Capabilities of an Argon Fluoride 193 nm excimer laser for laser ablation inductively coupled plasma mass spectrometry microanalysis of geological materials [J].
Gunther, D ;
Frischknecht, R ;
Heinrich, CA ;
Kahlert, HJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (09) :939-944
[9]   Laser ablation and arc/spark solid sample introduction into inductively coupled plasma mass spectrometers [J].
Günther, D ;
Jackson, SE ;
Longerich, HP .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1999, 54 (3-4) :381-409
[10]   NIST SRM 610, 611 and SRM 612, 613 multi-element glasses: Constraints from element abundance ratios measured by microprobe techniques [J].
Hinton, RW .