Mechanical performance of standard and cannulated 4.0-mm cancellous bone screws

被引:71
作者
Brown, GA
McCarthy, T
Bourgeault, CA
Callahan, DJ
机构
[1] Univ Minnesota, Dept Orthopaed Surg, Minneapolis, MN 55455 USA
[2] Hennepin Cty Med Ctr, Orthopaed Biomech Lab, Minneapolis, MN 55415 USA
[3] Minneapolis Med Res Fdn, Midwest Orthopaed Res Fdn, Minneapolis, MN USA
[4] Orthopaed Surg Consultants, Minneapolis, MN USA
关键词
D O I
10.1002/jor.1100180220
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The mechanical performance of bone screws is determined by their pull-out strength (holding power), compressive force, stripping torque, yield bending moment, ultimate bending moment, and fatigue strength, These parameters are related to the parameters of the screw design, including major thread diameter, minor thread diameter, thread length, pitch, shaft diameter, cannulation diameter, and material properties. The goal of the study was to theoretically predict the static performance of five 4.0-mm, 45-46-mm-long, cancellous, partially threaded standard and cannulated bone screws and compare the predictions with experimental measurements. A secondary goal was to determine if cannulation of the bone screw diminished its mechanical performance. The predicted values for pull-out force, compressive force, and stripping torque were determined by the thread length, major thread diameter, and thread shape factor. The screws with the largest major thread diameter and longest thread length had the greatest pull-out force, compressive strength, and stripping torque. However, when correcting for the thread length, a higher thread shape factor compensated for a smaller major diameter. The coefficient of determination (r(2)) for the correlation between the predicted and measured pull-out force improved from 0.75 to 0.90 when the theoretical model included the thread shape factor. The yield and ultimate bending moments are a function of the section modulus and material properties of the screw. The Ace solid screw had the greatest section modulus and yield and ultimate bending moments. The experimental data support the theoretical models for predicting the mechanical performance of bone screws. The design of the bone screws can be optimized on the basis of theoretical modeling. The strong correlation between the predicted and measured parameters allows comparison between bone screws without repeated experimental tests. Theoretical and experimental results show that cannulation of the bone screw did not inherently diminish its mechanical performance.
引用
收藏
页码:307 / 312
页数:6
相关论文
共 27 条
[1]
Ansell R H, 1968, J Biomech, V1, P279, DOI 10.1016/0021-9290(68)90023-7
[2]
Cancellous bone screw thread design and holding power [J].
Asnis, SE ;
Ernberg, JJ ;
Bostrom, MPG ;
Wright, TM ;
Harrington, RM ;
Tencer, A ;
Peterson, M .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1996, 10 (07) :462-469
[3]
*ASTM, 1983, ANN BOOK ASTM STAND, P106
[4]
BAKER BE, 1988, T ORTHOP RES SOC, V13, P447
[5]
IMPLANT HOLDING POWER OF THE FEMORAL-HEAD - A CADAVER STUDY OF FRACTURE SCREWS [J].
BENTERUD, JG ;
HUSBY, T ;
GRAADAHL, O ;
ALHO, A .
ACTA ORTHOPAEDICA SCANDINAVICA, 1992, 63 (01) :47-49
[6]
Factors affecting the pullout strength of cancellous bone screws [J].
Chapman, JR ;
Harrington, RM ;
Lee, KM ;
Anderson, PA ;
Tencer, AF ;
Kowalski, D .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (03) :391-398
[7]
DeCoster T A, 1990, J Orthop Trauma, V4, P169, DOI 10.1097/00005131-199004020-00012
[8]
DESIGN AND TESTING OF EXTERNAL FIXATOR BONE SCREWS [J].
EVANS, M ;
SPENCER, M ;
WANG, Q ;
WHITE, SH ;
CUNNINGHAM, JL .
JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (06) :457-462
[9]
FINLAY JB, 1989, CLIN ORTHOP RELAT R, P220
[10]
Hearn T C, 1993, J Orthop Trauma, V7, P138, DOI 10.1097/00005131-199304000-00006