Subharmonic emissions from microbubbles: Effect of the driving pulse shape

被引:27
作者
Biagi, Elena [1 ]
Breschi, Luca [1 ]
Varmacci, Enrico [1 ]
Masotti, Leonardo [1 ]
机构
[1] Univ Florence, Florence, Italy
关键词
D O I
10.1109/TUFFC.2006.157
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The aims of this work are to investigate the response of the ultrasonic contrast agents (UCA) insonified by different arbitrary-shaped pulses at different acoustic pressures and concentration of the contrast agent focusing on subharmonic emission. A transmission setup was developed in order to insonify the contrast agent contained in a measurement chamber. The transmitted ultrasonic signals were generated by an arbitrary wave generator connected to a linear power amplifier able to drive a single-element transducer. The transmitted ultrasonic pulses that passed through the contrast agent-filled chamber were received by a second transducer or a hydrophone aligned with the first one. The radio frequency (RF) signals were acquired by fast echographic multiparameters multi-image novel apparatus (FEMMINA), which is an echographic platform able to acquire ultrasonic signals in a real-time modality. Three sets of ultrasonic signals were devised in order to evaluate subharmonic response of the contrast agent respect with sinusoidal burst signals used as reference pulses. A decreasing up to 30 dB in subharmonic response was detected for a Gaussian-shaped pulse; differences in subharmonic emission up to 21 dB were detected for a composite pulse (two-tone burst) for different acoustic pressures and concentrations. Results from this experimentation demonstrated that the transmitted pulse shape strongly affects subharmonic emission in spite of a second harmonic one. In particular, the smoothness of the initial portion of the shaped pulses can inhibit subharmonic generation from the contrast agents respect with a reference sinusoidal burst signal. It also was shown that subharmonic generation is influenced by the amplitude and the concentration of the contrast agent for each set of the shaped pulses. Subharmonic emissions that derive from a nonlinear mechanism involving nonlinear coupling among different oscillation modes are strongly affected by the shape of the ultrasonic driving pulse.
引用
收藏
页码:2174 / 2182
页数:9
相关论文
共 23 条
[1]  
BIAGI E, 2003, 27 INT S AC IM SAARB
[2]   Nondestructive subharmonic imaging [J].
Chomas, J ;
Dayton, P ;
May, D ;
Ferrara, K .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2002, 49 (07) :883-892
[3]  
DAYTON P, 1999, IEEE T ULTRASON FERR, V46, P1
[4]   GENERATION OF SUBHARMONICS OF ORDER ONE-HALF BY BUBBLES IN A SOUND FIELD [J].
ELLER, A ;
FLYNN, HG .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1969, 46 (3P2) :722-&
[5]  
Forsberg F, 2004, ULTRASON, P5
[6]   In vivo pressure estimation using subharmonic contrast microbubble signals: Proof of concept [J].
Forsberg, F ;
Liu, JB ;
Shi, WT ;
Furuse, J ;
Shimizu, M ;
Goldberg, BB .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2005, 52 (04) :581-583
[7]   Subharmonic imaging of contrast agents [J].
Forsberg, F ;
Shi, WT ;
Goldberg, BB .
ULTRASONICS, 2000, 38 (1-8) :93-98
[8]   High frequency nonlinear B-scan imaging of microbubble contrast agents [J].
Goertz, DE ;
Cherin, E ;
Needles, A ;
Karshafian, R ;
Brown, AS ;
Burns, PN ;
Foster, FS .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2005, 52 (01) :65-79
[9]   Subharmonic generation from ultrasonic contrast agents [J].
Krishna, PD ;
Shankar, PM ;
Newhouse, VL .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (03) :681-694
[10]   NUMERICAL INVESTIGATION OF NONLINEAR OSCILLATIONS OF GAS-BUBBLES IN LIQUIDS [J].
LAUTERBORN, W .